232 Section VI ■ Respiratory Care
b. Velcro (to make attachment circles and moustache
c. DuoDERM (to make nasal septum protective layer)
E. Technique (See Procedures Website
Nonventilator-derived b-CPAP apparatus involves
making a simple water seal device that can be put
together in neonatal units. It consists of a container of
water, through which the expiratory gas from the baby
is bubbled at a measured level below the surface (e.g.,
5 cm below the surface = 5cm H2O CPAP). The lower
the level of the tip of the expiratory tubing below the
surface of the water, the higher the CPAP (Fig. 35.1). It
is important to fix the water bottle to an IV pole at or
a. Before attaching the device to an infant
(1) Position the infant with the head of the bed elevated 30 degrees.
(2) Gently suction the mouth, nose, and pharynx.
Whenever possible, use size 8-Fr suction
catheter. Smaller-sized catheters are not as efficient.
(3) Place a small roll under the infant’s neck/shoulder.
Allow slight neck extension to help maintaining
(4) Clean the infant’s upper lip with water.
(5) Place a thin strip of DuoDERM (or Tegaderm)
over the upper lip. That should also cover the
nasal columella and both sides of nasal apertures (Fig. 35.2).
(6) Cut a Velcro moustache and fix it over the
(7) Cut two strips of soft Velcro (8 mm width) and
wrap them around the transverse arm of the
device, about 1 cm away, on each side, from the
b. Placing nasal prongs into infant’s nostrils (Figs. 35.2
(1) Use appropriate-size CPAP prongs. The correct
size nasal prongs should snugly fit the infant’s
nares without pinching the septum. If prongs
If prongs are too large, they may cause mucosal
(2) Curve prongs gently down into the infant’s
(3) Press gently on the prong device until the soft
Velcro strips adhere to the moustache.
(4) Make sure of the following points
(a) Nasal prongs fit well in the nostrils
(b) Skin of nares is not stretched (indicated by
blanching of the rim of the nostrils)
(c) Corrugated tubes are not touching the
(d) There is no lateral pressure on the nasal septum
(e) There is a small space between the nasal
septum and the bridge between the prongs
(f) Prongs are not resting on the philtrum
Fig. 35.1. Bubble CPAP circuit. This simplified
diagram demonstrates the components of the b-CPAP
infant from the wall source after it is warmed and
humidified. The free expiratory limb of the tube is
immersed under the surface of sterile water to produce
the required CPAP (usually 5 cm H2O).
No comments:
Post a Comment
اكتب تعليق حول الموضوع