(5) After 1 minute, press Temp Set (Fig. 45.18D).
(6) Press the ∆/∇ (Fig. 45.18) button to set the
The status display (Fig. 45.19C) will read 33.5°C.
(7) Press Gradient Variable (Fig. 45.19H).
(8) Press ∆/∇ (Fig. 45.16) to 20°C.
(9) Press the Gradient Variable (Fig. 45.19H) again.
(10) Listen for the activation of the pump.
(11) Check that the water flow indicator (Fig.
(12) Place the dry sheet or disposable cover over the
infant, to decrease convection losses and fluctuations in water temperature in blanket (Fig.
(13) Monitor temperature displays. The Patient display (Fig. 45.19E) will show the infant’s actual
The Water display (Fig. 45.19A) will show
the actual temperature of the circulating
The Status display (Fig. 45.19C) will show
the mode of operation and set temperature.
Monitor core temperature every 15 minutes
to determine when the target temperature is
Maintain HT (Gradient Variable Mode)
1. Press Temp Set (Fig. 45.19D).
2. Press the ∆/∇ (Fig. 45.19) button to maintain core
3. Press Gradient Variable (Fig. 45.19H).
4. Press the ∆/∇ (Fig. 45.19) button to 5°C to minimize
the temperature fluctuations between the patient and
the water in the cooling blanket.
5. Press the Gradient Variable (Fig. 45.19H) again.
6. Listen for the activation of the pump.
7. Check that the water flow indicator (Fig. 45.18F) is
Initiate Manual Rewarming after 72 hours of HT
1. Press Temp Set (Fig. 45.19D).
3. Increase 0.5°C every hour until the core temperature is
4. Press Gradient Variable (Fig. 45.19H).
5. Press ∆ to 5°C to minimize the temperature fluctuations
between the patient and the water in the cooling blanket.
6. Press Gradient Variable (Fig. 45.19H).