which pass through the rest of the plate.
Supposing that the specimen is suspension free fluid,
the only light that reaches the eye is that which goes from
the annulus through the phase plate. Whereas presence
of organisms would diffract and scatter the light. The light
passing through the fluid gets out of phase with the light that
has the organisms stand out in contrast to their background.
An annulus, a phase plate and a telescope that is needed
for adjusting the rings of both annulus and the phase plate.
1. Focus the specimen with the right objective after
2. Place the matching annulus at its position.
3. Remove the eyepiece and put the telescope in its place,
adjust it till the two rings, one bright and one dark are
4. Adjust condenser screws till the bright annulus ring
fits exactly into the darker ring of the phase plate.
5. Remove the telescope, replace the eyepiece, focus and
This method is made use of for examining live organisms,
It can also be used for platelet counting and for examining routine urine specimens.
a. A halo is seen around each particle, it gives a false
b. In addition, some resolution power is lost but this is
more than compensated for by the increased contrast
This method too, is used for visualizing organisms
suspended in fluid, both the structure and the motility
FIG. 1.10: Working of an oil immersion objective
of the organisms can be seen. In this method, the light
enter the objective. Instead the light is reflected to pass
through the outer rim of the condenser at a wide angle
which illuminates the microorganisms by a ring of light
In this method, the light that is seen comes only from
the microorganisms themselves and not from the light
source. Hence, the organisms are brightly illuminated
against a dark background. Though useful, this method is
1. An oil immersion dark ground condenser with the
2. A funnel stop for insertion in 100X objective to reduce
its NA and exclude light coming directly from the
3. A brightly illuminated microscope lamp.
4. Scratchless slides not more than 1 mm thick.
1. Fit the dark ground condenser and raise it to stage level.
2. Place the coverslipped specimen on the thin polished
glass slide. Both, the coverslip and the slide should be
3. Place a drop of immersion oil between the condenser
4. Adjust light source and the mirror properly.
5. Focus 10X objective and observe.
6. Focus condenser up or low, so that the ring ultimately
No comments:
Post a Comment
اكتب تعليق حول الموضوع