4. Reagent red cells may be stored as whole blood with

CPD or ACD but more usually they are stored as

washed cells in a preservative solution. A modified

Alsever’s solution, with added inosine and with

antibiotics is commonly used permitting satisfactory

storage for at least 35 days at 4°C.

¾ Clotted whole blood should be tested within 14 days

¾ Anticoagulated blood using various anticoagulants

should be tested within specific time limits as follows:

EDTA or heparin: 2 days

Sodium citrate or sodium oxalate: 14 days

ACD or CPD: 28 days.

Equipment

¾ Prior to testing, check whether pipettes and other

instruments used are well calibrated and in proper

working condition.

Reagents

¾ The blood grouping reagents contain 0.1% sodium

azide as preservative. Avoid contact of the reagent

with skin and mucosa. On disposal flush with large

quantities of water

¾ Adhere to appropriate storage conditions of the reagent

as mentioned in the respective package inserts.

Common Causes of False Negative and

False Positive Results in ABO Testing

False Negative Results

¾ Reagent or test serum not added to a tube

¾ Hemolysis not identified as a positive reaction

¾ Inappropriate ratio of serum (or reagent) to red cells

¾ Tests not centrifuged sufficiently

¾ Tests incubated at temperatures above 20–24°C

¾ Incorrect interpretation of test results.

False Positive Results

¾ Overcentrifugation of tubes

¾ Use of contaminated reagents, red cells or saline

¾ Use of dirty glassware

¾ Incorrect interpretation of results.

Interpretation of Agglutination Reactions

To avoid false readings and to standardize recording of

results, tests should be read with the aid of the microscope

(Table 11.10).

TABLE 11.10: Interpretation of agglutination reactions

++++ One complete mass of agglutination, readily visible on the

slide before microscopic examination

+++ Large separate masses of agglutination, readily visible on

the slide before microscopic examination

++ Smaller agglutinates, still readily visible on the slide

before examination under the microscope

+ A granular appearance, just visible on the slide before

examination under the microscope. The microscope

reveals big clumps of agglutinates of more than 20 cells

(+) Smaller clumps (12–20) cells only detectable by use of

microscope

GW Good weak—clumps of 8–12 cells only detectable

microscopically

W Weak—Weak reactions with uniform distribution of small

clumps of 4–6 cells

? Sticky—uneven distribution of cells, 2 or 3 sticking

together, more noticeable when the cells are “rolling” on

the slide

0 or - All cells free and evenly distributed. The 0 sign is

preferred since—can easily be altered to +

MF/NF Mixed filed or negative field—agglutinates in a field of free

cells, e.g. the type of reaction observed with mixed bloods

374 Concise Book of Medical Laboratory Technology: Methods and Interpretations ABO Grouping

Problem: False positive results

Possible causes Solutions

1. Bacterial contamination (some bacteria such as

Staphylococcus aureus will agglutinate all red cell

samples thereby giving false positive results)

Check the reagents for turbidity. Extreme turbidity may indicate bacterial

contamination. Such contaminated reagents should not be used for testing

2. The antisera may be contaminated with another

antibody. There may have been accidental mixing of

antisera; this could be due to the interchanging of

caps. For example, if either anti-A, anti-B or anti-A, B is

contaminated with anti-D as anti-D reagent is colorless

and contamination is not visually observed in such a

case the anti A or anti B or anti A, B reagent will give

false positive results with all Rh positive blood groups

Check the color of the cap on the antisera bottle. Also check the antisera

with the blood of a person whose blood group is known. If anti-A, anti-B

or anti-A, B vial is contaminated with anti D, then the anti-A, anti-B or

anti-A, B should be checked with Rh negative blood sample, if it gives no

agglutination thereby confirming anti-D contamination or else it

could be some other contamination

3. Particles of dust, debris, chemicals or detergents

on the slide or in the tube giving non-specific

agglutination

Clean and dry glassware should be used while carrying out the test

4. Peripheral drying or fibrin strands were mistaken for

agglutination in case of slide test

The test should not be carried out directly under the fan. The vials should be

capped immediately after use. The results of the test should be read at

2 minutes not beyond as drying may be interpreted as positive result. Except

in case of anti-A1 lectin, the agglutination should be observed at one minute

5. Excessive centrifugation in case of tube test Ensure that centrifugation is carried out for either one minute at 1000 rpm

or 20 seconds at 3400 rpm. Centrifugation should be adequate to produce

a cell button with a clear supernatant but without packing the cells so

tightly that they are difficult to dislodge. Each laboratory must interval its

equipment at regular interval

Problem: False negative results

Possible causes Solutions

1. Storage of antisera at higher/lower temperatures than

specified

Reagents should be stored at 2–8°C when not in use. Thermal damage due

to faulty storage may result in a loss of reactivity. Check the turbidity of the

reagents; extreme turbidity may also be due to thermal damage. Antibody

activity decreases at lower temperatures. Do not freeze the reagents

2. Blood sample stored for too long Check the period of time for which the blood sample has been sotred.

Anticoagulated blood using various anticoagulants should be tested within

the below mentioned time period (when stored at 2–8°C)

• EDTA or heparin—2 days

• Sodium oxalate or sodium citrate—14 days

• ACD or CPD—28 days

• Clotted whole blood—14 days

3. Use of hemolyzed samples Check the samples before use. Do not use hemolyzed samples

4. Prozoning (zone of antibody excess) or postzoning

(zone of antigen excess)

The antigen and the antibody should be present in optimal concentrations

of the agglutination to be seen properly. Check the sample volume and

the reagent volume used. Both the sample and reagent volume should be

equal in slide test and a 5% suspension of cells should be used in tube test.

Ensure that there are no air bubbles while dispensing samples and reagents

5. Outdated or contaminated reagent Check the expiry date of the reagent. Also check the reagent for

contamination. Extreme turbidity may be due to microbial contamination

Contd...

Blood Banking (Immunohematology) 375

Possible causes Solutions

6. Blocking effect when the cells are coated, e.g. in

hemolytic disease of the newborn (HDN), acquired

immune hemolytic anemia (IHA)

Check the history of the patient for the presence of disorders in which case

the cells may be coated in vivo. This can be established by doing a DAT on

washed cells

7. Weak antigens In the tube test, every tube with negative reaction should be centrifuged and

the results should be read again after 5 minutes, so that weak antigens are

not overlooked

8. Under centrifugation Centrifugation should be carried out for 1 minute at 1000 rpm or for 20

seconds at 3400 rpm as per recommended procedures to give enough time

for antigen-antibody to bind

9. Vigorous shaking for resuspension of cells after

centrifugation

Resuspend the cells slowly and gently after centrifugation. Each laboratory

must calibrate its equipment at regular intervals

10. In case of A1 lectin, A antigen is not fully expressed on

the red cells of newborns below one year of age

Check the age of the patient. If below 1 year repeat typing after child reaches

1 year of age

Problem: Hemolysis or red blood cells

Possible causes Solutions

1. Use of wet slides and tubes Wet glassware can cause hemolysis of RBC’s. Ensure that only dry

glassware is used for testing

Problem: Weak agglutination

Comments

Search This Blog

Archive

Show more

Popular posts from this blog

TRIPASS XR تري باس

CELEPHI 200 MG, Gélule

ZENOXIA 15 MG, Comprimé

VOXCIB 200 MG, Gélule

Kana Brax Laberax

فومي كايند

بعض الادويه نجد رموز عليها مثل IR ، MR, XR, CR, SR , DS ماذا تعني هذه الرموز

NIFLURIL 700 MG, Suppositoire adulte

Antifongiques مضادات الفطريات

Popular posts from this blog

علاقة البيبي بالفراولة بالالفا فيتو بروتين

التغيرات الخمس التي تحدث للجسم عند المشي

إحصائيات سنة 2020 | تعداد سكَان دول إفريقيا تنازليا :

ما هو الليمونير للأسنان ؟

ACUPAN 20 MG, Solution injectable

CELEPHI 200 MG, Gélule

الام الظهر

VOXCIB 200 MG, Gélule

ميبستان

Popular posts from this blog

TRIPASS XR تري باس

CELEPHI 200 MG, Gélule

Popular posts from this blog

TRIPASS XR تري باس

CELEPHI 200 MG, Gélule

ZENOXIA 15 MG, Comprimé

VOXCIB 200 MG, Gélule

Kana Brax Laberax

فومي كايند

بعض الادويه نجد رموز عليها مثل IR ، MR, XR, CR, SR , DS ماذا تعني هذه الرموز

NIFLURIL 700 MG, Suppositoire adulte

Antifongiques مضادات الفطريات

Popular posts from this blog

Kana Brax Laberax

TRIPASS XR تري باس

PARANTAL 100 MG, Suppositoire بارانتال 100 مجم تحاميل

الكبد الدهني Fatty Liver

الم اسفل الظهر (الحاد) الذي يظهر بشكل مفاجئ bal-agrisi

SEDALGIC 37.5 MG / 325 MG, Comprimé pelliculé [P] سيدالجيك 37.5 مجم / 325 مجم ، قرص مغلف [P]

نمـو الدمـاغ والتطـور العقـلي لـدى الطفـل

CELEPHI 200 MG, Gélule

أخطر أنواع المخدرات فى العالم و الشرق الاوسط

Archive

Show more