but in most instruments the intensity is below the
danger level. Methods in which photochemical reactions
are likely to occur usually mention the precautions to be
dark bottle, because on exposure to light a photochemical
reaction takes place and the reagent gets oxidised and
In some colorimetric reactions, the color may be stable for
only a short period of time. It is then necessary to time the
reaction carefully so that the readings of all samples and
standards are made during the time that the color remains
constant. Instability of color may be due to temperature
or absorption by the walls of the container where these
factors have an influence, they must be kept constant for
d. Foreign Matter and Air Bubbles
Solutions in the cuvettes must be free of lint or other
foreign matter, and air bubbles. A scrupulous cleaning
of the cuvettes and other glassware used in the analysis
should help to eliminate foreign matter from the solution.
to a decrease in optical density of the solution.
e. Errors of Weighing and Dilution
Simple errors of weighing and dilution in preparing
reagents, sample and standards can affect photometric
results appreciably. A good analytical balance and reliable
volumetric glassware should be used. For example, while
reconstituting a control serum care should be taken while
dispensing the volume of distilled water to the control
sera bulb. An error in dilution will result in change in the
concentration of the constituents.
Instruments are capable of considerable precision of
measurement. The instrumental precision is of a higher
order than that normally resulting from the development
of color in the test solutions. Inherent errors of the solution
as mentioned above actually cause greater deviation than
Unless a double cell photometer is being used, the
consistency and reproducibility of the light source is
important. Fluctuations in voltage should be overcome by
the use of a voltage stabilizer in line-operated instruments.
Lamps should be allowed to warm up for at least 5 minutes
before steady output can be expected.
Stray light from windows or overhead lighting striking
the instrument can cause error since invisible particles
suspended in solutions can reflect these rays. The covering
of the cuvette compartment with a light-tight cover (as
usually provided with the instrument) before taking
readings will reduce this error.
In the prism spectrophotometer, the purity of the
monochromatic band depends on the width of the
entrance and exit slits. The use of a narrow slit width will
produce more accurate results.
Moisture can be the cause of fluctuating readings in
spectrophotometric operation. In instruments employing
desicant, it is advisable to change the freshly dried silica
gel at regular intervals. This is particularly important in an
e. Linearity of Photocell Response
Reliable results depend upon the current output of
the photocell being proportional to the light striking
the photocell. This relationship can be disturbed if the
photocell is not adequately protected from moisture and
from overheating. Some instruments are fitted with heat
absorbing filters in the optical system and provided with
thermal insulation of the light source.
An important source of error and one which requires
constant checking is the cuvette. It is necessary that cuvettes
be optically matched, so that readings will not be influenced
by their individual variation when a series is used for making
measurements. When it is necessary to reuse cuvettes that
have not had adequate time to dry after cleaning, a rinse
with alcohol and ether or acetone may be used to speed up
the drying process. If cuvettes are dirty, etched, scratched
or marked with fingerprints, erroneous readings will result.
No comments:
Post a Comment
اكتب تعليق حول الموضوع