Possible causes Solutions

1. Preleukemic states and acute myeloblastic leukemia Check the clinical history of the patient

2. Acquired loss of antigens occur in healthy elderly

individuals

Check the age of the patient

Problem: Mixed field agglutination in ABO grouping or discrepancy observed between the red cell group

and the reverse group

Possible causes Solutions

1. Transfusion of blood of a different group. (e.g. group O

to group A or more significantly group A to group O)

It is extremely important to verify the cause of mixed field agglutination. An

investigation of the patient’s history may help to verify the cause of mixed

field agglutination

2. Chimera (this is a condition where an individual

possesses more than one population of red cells)

This may be caused by exchange of blood cells during early fetal life of

nonidentical twins or may be artificially induced by compatible but not same

blood group. Check the history of the patient

3. Blood group antigens altered by bacterial enzymes.

Some group A individuals with intestinal obstruction,

carcinoma of the colon or rectum and other disorders

of the lower intestine acquire a B-like antigen

Check the history of the patient for the presence of any of these conditions

and perform reverse grouping in case of acquired B to confirm the blood

group

Problem: Delayed agglutination

Possible causes Solutions

1. Subgroups of A, other than A1 such as A2, A3, Ax, etc.

have weakly expressed A antigenic sites on the red

cells, hence these red cells give weak reaction with

anti-A reagent as compared to A1 cells

Check these cells with Anti-A1 lectin, these cells should not react with anti-A1

lectin

Graded reaction helps the clinician to differentiate between strong (A1) and

other weaker subgroups of A group

2. Reagents used immediately after removing from the

refrigerator

The reagent vial must be brought to room temperature prior to starting the

test

3. The antigen and antibody are not present in optimal

concentrations

The sample volume and the reagent volume dispensed should be as per the

instructions given in the protocol

Contd...

376 Concise Book of Medical Laboratory Technology: Methods and Interpretations Rh Typing

Problem: False positive results

Possible causes Solutions

1. Bacterial contamination (some bacteria will

agglutinate all red cell samples thereby giving

false positive results)

Check the reagents for turbidity. Extreme turbidity may indicate bacterial

contamination. Such contaminated reagents should not be used for testing

2. The antisera may be contaminated with

another antibody. There may have been

accidental mixing of antisera; this could be

due to interchanging of caps

Check the color of the cap on the antisera bottle. Also check the antisera with the blood

of a person whose blood group is known

3. Particles of dust, debris, chemicals or

detergents on the slide or in the tube giving

nonspecific agglutination

Clean and dry glassware should be used while carrying out the test

4. Peripheral drying or fibrin strands were

mistaken for agglutination in case of slide test

The test should not be carried out directly under the fan. The results should be read

within 2 minutes. The vial should be capped immediately after use

5. Excessive centrifugation in case of tube test Ensure that centrifugation is carried out for either one-minute at 1000 rpm or 20

seconds at 3400 rpm. Centrifugation should be adequate to produce a cell button with

a clear supernatant but without packing the cells so tightly that they are difficult to

dislodge. Each laboratory must calibrate its equipment at regular intervals

6. Rouleaux formation in case of tube test may

be mistaken for agglutination. Rouleaux

formation is said to occur when the red

cells appear like a stack of coins. Rouleaux

formation may be caused by the following

Rouleaux formation can be distinguished from agglutination by adding 2 drops of

saline to the reaction mixture; if the clumping of cells dissolves then it indicates

rouleaux formation

Check the history of the patient for the conditions like multiple myeloma.

The suspension of the cells prepared should be as per the instructions in the package

insert

Problem: False negative results

Possible causes Solutions

1. Storage of antiseras at higher/lower

temperatures than specified

Reagents should be stored at 2–8°C when non in use. Thermal damage due to faulty

storage may result in a loss of reactivity. Check the turbidity of the reagents; extreme

turbidity may be due to thermal damage. Do not freeze the reagents

2. Blood sample stored for too long than

recommended

Check the period of time for which the blood sample has been stored. Anticoagulated

blood using various anticoagulants should be tested within the below mentioned time

period (when stored at 2–8°C)

• EDTA or heparin—2 days

• Sodium oxalate or sodium citrate—14 days

• ACD or CPD—28 days

• Clotted whole blood—14 days

3. Use of hemolyzed samples Check the samples before use. Do not sue hemolyzed samples.

4. Prozoning (zone of antibody excess) or

postzoning (zone of antigen excess)

The antigen and the antibody should be present in optimal concentrations of the

agglutination to be seen properly. Check the sample volume and the reagent volume

used. Both the sample and reagent volume should be equal in slide test and a 5%

suspension of cells should be used in tube test. Ensure that there are no air bubbles

while dispensing samples and reagents

5. Outdated or contaminated reagent Check the expiry date of the reagent. Also check the reagent for contamination.

Extreme turbidity may be due to microbial contamination

6. Blocking effect when the cells are coated, e.g.

in hemolytic disease of the newborn (HDN),

acquired immune hemolytic anemia (IHA)

Check the history of the patient of the presence of disorders in which the cells may be

coated

Contd...

Blood Banking (Immunohematology) 377

Possible causes Solutions

7. Weak antigens In the tube test, every tube with negative reaction should be centrifuged and results

should be read again after 5 minutes, so that weak antigens are not overlooked

8. Under centrifugation Centrifugation should be carried out for 1 minute at 1000 rpm or for 20 seconds at

Comments

Search This Blog

Archive

Show more

Popular posts from this blog

TRIPASS XR تري باس

CELEPHI 200 MG, Gélule

ZENOXIA 15 MG, Comprimé

VOXCIB 200 MG, Gélule

Kana Brax Laberax

فومي كايند

بعض الادويه نجد رموز عليها مثل IR ، MR, XR, CR, SR , DS ماذا تعني هذه الرموز

NIFLURIL 700 MG, Suppositoire adulte

Antifongiques مضادات الفطريات

Popular posts from this blog

علاقة البيبي بالفراولة بالالفا فيتو بروتين

التغيرات الخمس التي تحدث للجسم عند المشي

إحصائيات سنة 2020 | تعداد سكَان دول إفريقيا تنازليا :

ما هو الليمونير للأسنان ؟

ACUPAN 20 MG, Solution injectable

CELEPHI 200 MG, Gélule

الام الظهر

VOXCIB 200 MG, Gélule

ميبستان

Popular posts from this blog

TRIPASS XR تري باس

CELEPHI 200 MG, Gélule

Popular posts from this blog

TRIPASS XR تري باس

CELEPHI 200 MG, Gélule

ZENOXIA 15 MG, Comprimé

VOXCIB 200 MG, Gélule

Kana Brax Laberax

فومي كايند

بعض الادويه نجد رموز عليها مثل IR ، MR, XR, CR, SR , DS ماذا تعني هذه الرموز

NIFLURIL 700 MG, Suppositoire adulte

Antifongiques مضادات الفطريات

Popular posts from this blog

Kana Brax Laberax

TRIPASS XR تري باس

PARANTAL 100 MG, Suppositoire بارانتال 100 مجم تحاميل

الكبد الدهني Fatty Liver

الم اسفل الظهر (الحاد) الذي يظهر بشكل مفاجئ bal-agrisi

SEDALGIC 37.5 MG / 325 MG, Comprimé pelliculé [P] سيدالجيك 37.5 مجم / 325 مجم ، قرص مغلف [P]

نمـو الدمـاغ والتطـور العقـلي لـدى الطفـل

CELEPHI 200 MG, Gélule

أخطر أنواع المخدرات فى العالم و الشرق الاوسط

Archive

Show more