5. Test results are correct but reporting may be wrong Check and verify final report correctly
6. Test interpretations are wrong Results should be interpreted as per test protocol
7. Drying of the reagent on slide Do not read results beyond 2 minutes
Do not perform the test directly under a fan
8. Male urine used as negative
control to check the working of
Do not use male urine as negative control as it will give false positive results
Since the hormones, protein and salt composition of male urineis different
and foresight being a direct test system is calibrated to yield accurate results
on male urine. Use the negative control provided with the kit or use known
Problem: False Negative Results
1. Wrong sample used Collect, label the samples appropriately and test accordingly
2. Sample stored for a long period of time 0.01%
should be used as preservatives. The samples can
Specimens should be tested preferably within 12 hours of collection
Should a delay in testing occur, thimerosal 0.001% or sodium azide. at 2–8°C
3. Reagent may have deteriorated due to thermal
Avoid exposure of the reagents to high temperatures. Store at 2–8°C. Do not
4. Prozoning due to hCG levels above 250 IU/mL (In
pathological conditions such as choriocarcinoma
In such cases, the sample should be diluted and used. Check the patient’s
5. Serum is being used as specimen Use only urine and not serum as specimen.
Ensure that there are no air bubbles while dispensing the sample.
7. Reagents not brought to room temperature before
Bring all reagents to room temperature before testing.
8. Presence of detergent on the slide Wash thoroughly with water and retest
Problem: False Positive Results
1. Presence of trophoblastic and non-trophoblastic neoplasms
such as choriocarcinoma and hydatidiform mole
Check the clinical history of the patient
2. The flow properties of the nitrocellulose membrane are partially
affected leading to the nonspecific movement of partially
Check the device/dipstic pouch for pinholes and observe the
dessicant for any color change. The results of the test should be
correlated with clinical findings
Problem: Delayed Positive Results
1. Urine samples were tested immediately after removing from
Bring the urine samples to room temperature before testing
Problem: False Negative Results
2. The kit is exposed to very high temperatures leading to the
deterioration of the antibodies coated on the device/dipstick
Store the kit at 4–30°C when not in use
testing urine samples can be stored at 2–8°C up to 72 hours. Turbid
samples should be centrifuged before testing
4. Highly diluted samples or samples of very early pregnancy used
Highly diluted samples or samples of very early pregnancy may not
contain representative levels of hCG. In such cases, if pregnancy is
suspected, the test should be repeated with first morning urine after
48–72 hours after the initial test
1. Pinholes or defect in the pouch. The nitrocellulose membrane
has lost its flow properties due to the absorbance of moisture
Check the pouch for pinholes and also check the color of the
dessicant (silica gel) accompanying the pouch
A change in color from deep blue to white/pink indicates absorbance
of moisture. In such cases, discard the test device and rerun the test
2. The device/dipstick is removed from the refrigerator and tested
immediately before attaining room temperature. This leads
to the hydration of the sites on the nitrocellulose membrane
thereby adversely affecting its flow properties
The test device/dipstick should be brought to room temperature
1. A number of other conditions including trophoblastic
and non-trophoblastic neoplasms such as hydatidiform
mole, choriocarcinoma cause elevated levels of hCG
Check the clinical history of the patient, before diagnosing for pregnancy
2. The flow properties of the nitrocellulose membrane are
partially affected leading to the nonspecific movement of
partially aggregated gold-sol particles
Check the device/dipstick pouch for pinholes besides also note color
No comments:
Post a Comment
اكتب تعليق حول الموضوع