Plasma bicarbonate ↑ 10–15% Compensation for chronic hyperventilation
Urine glucose Mild glycosuria with normal blood sugar levels Lowered renal threshold
Renal function GFR increases, Concentrating capacity ↑ Mobilization of retained fluid
Free T4 ↑ and normal respectively
Total and free T3 ↑ and normal respectively
ACTH Markedly ↑ in early months ?
Serum total cortisol ↑ Estrogen—associated rise in binding protein
Serum testosterone ↑ ? ↑ ovarian synthesis
Aldosterone Progressive ↑ Altered plasma volume, renal sodium load
Renin—angiotensin ↑ in latter half With ↑ Na or ↑ blood pressure, level decreases
Test Procedure and Interpretation of Results
1. Bring the sealed pouch to room temperature, open the
pouch and remove the device. Once opened, the device
2. Dispense two drops of urine/serum specimen into
the sample well ’S’ using the dropper provided.
Refrigerated specimens must be brought to room
Read the results at the end of 5 minutes for urine
samples and at the end of 15 minutes for serum
Negative: Only one colored-band
appears on the control region ‘C’.
Positive: In addition to the control band, a distinct colored-band
also appears on the test region.
3. The test should be considered invalid if no band
appears. Repeat the test with a new device.
4. Although, depending on the concentration of hCG
in the specimen, positive results may start appearing
as early as 30 to 60 seconds, negative results must be
confirmed only at the end of the stipulated time.
Indirect/Latex Agglutination Inhibition Method
Problem: False Positive Results
1. Latex reagent not working The reagents should be stored at 2–8°C. Do not freeze the reagents
Check the working of the latex reagent by mixing with the anti-β hCG reagent
No agglutination indicates deterioration of the latex reagent
2. Anti-β hCG reagent vial is contaminated with known
positive sample. This may be due to the dropper tip
Check the anti-β hCG reagent with the latex reagent. If no agglutination is
observed, it indicates that the anti-β hCG reagent is not working
Vial droppers must not be interchanged and reagents of the same lot numbers
Collect urine specimen in a detergent free container
4. Very high hCG levels are found in pathological
conditions like choriocarcinoma and hydatidiform
Do a semiquantitative test and check for high abnormal hCG levels
Problem: Delayed Agglutination
1. Borderline case with hCG levels just about
Use the first morning sample. Retest fresh sample after a week
3. Reagents not brought to room temperature before
All reagents must be brought to room temperature before testing. This is
necessary for the antigen-antibody to react optimally
4. Delayed testing Sample should be tested within 12 hours of collection
6. Too early diagnosis of hCG levels below detection
Retest after a week using 1st morning sample
Problem: False Negative Results
1. Case of ectopic pregnancy Follow-up with the patient’s history
2. Threatened abortion progressing into inevitable
Repeat test and correlate results with clinical findings
3. The urine sample is contaminated, cloudy and
bloody or is used after a long period of time
Specimen should be collected in clean glass of plastic containers free of
detergents or any such material
Specimen should be tested preferably within 12 hours of collection
They can be stored at 2–8°C for up to 72 hours after adding preservatives
thimerosal (0.001%) or sodium azide (0.01%). If the specimen is cloudy
or bloody, centrifuge it at 1000 rpm for one minute and use only the clear
Problem: False Positive Results
collected in a clean and dry container
Do not use grossly contaminated specimens. If the specimen is hazy or bloody,
centrifuge at 1000 rpm for one minute and use clear supernatant for testing
2. Wrong sample tested Label and test correct sample. Record results accordingly
3. Latex reagent contaminated with positive control/
Ensure that the dropper tip does not touch the sample/control on the slide
Take one or two drops of the latex reagent on a slide, and check for
granulation, aggregation or agglutination. There should be no granulation,
4. Presence of detergent on slide Wash slide thoroughly with water,
No comments:
Post a Comment
اكتب تعليق حول الموضوع