Human red blood cells are classified as RhoD positive or
RhoD negative depending on the presence or absence of
D antigen on them. Approximately, 85% of the Caucasian
population is Rh positive. The Du phenotype is a variant of
D antigen and is recognized by performing the antiglobulin
About 60% of the traditional Dus, now classified as
weak or partial D’s may react with Eryclone anti-D (IgM)
in slide tests and about 90% may be detected by the tube
Eryclone anti-D (IgM) is a ready-to-use reagent, prepared
from supernatants of cell cultures with antibody
producing B-lymphocytes obtained through EBV
transformation and is a blend of monoclonal antibodies
of immunoglobulin class IgM. These antibodies are a
mixture of several monoclonal antibodies of the same
specificity but having the capability of recognizing
different epitopes of human red blood cell antigen D
Eryclone Anti-D (IgM) does not detect all weak and
partial D’s. For the confirmation of negative reactions with
Eryclone Anti-D (IgM) further testing with an incomplete
Anti-D of IgG class is strongly recommended to confirm
the presence or absence of weak/partial D’s.
Each batch of reagent undergoes rigorous quality
control at various stages of manufacture for its specificity,
a. Store the reagent at 2–8°C. Do not freeze.
b. The shelf life of the reagent is as per the expiry date
mentioned on the reagent vial label.
Human red blood cells possessing D antigen will
agglutinate in the presence of antibody directed towards
the antigen. Agglutination of red blood cells with Eryclone
anti-D (IgM) reagent is a positive test result and indicates
the presence of D antigen. No agglutination with the
reagent is a negative test result and indicates the absence
of D antigen. All negative test results should be further
tested for D (presence of weak/partial Ds) by performing
the Du test procedure using an incomplete Anti-D of IgG
1. In vitro diagnostic reagent for laboratory and professional use only. Not for medicinal use.
2. Eryclone Anti-D (IgM) reagent is not from human
source, hence, contamination due to HBsAg and HIV
3. The reagent contains sodium azide 0.1% as preservative.
Avoid contact with skin and mucosa. On disposal flush
with large quantities of water.
4. Extreme turbidity may indicate microbial contamination or denaturation of protein due to thermal
damage. Such reagent should be discarded.
No special preparation of the patient is required prior
to sample collection by approved techniques. Samples
should be stored at 2–8° C if not tested immediately. Do
not use hemolyzed samples. Anticoagulated blood using
various anticoagulants should be tested within the below
Sodium citrate or sodium oxalate : 14 days
Clotted whole blood should be tested within 14 days.
Additional Material Required for Slide and Tube
Glass slides (50 × 75 mm), test tubes (10 × 75 mm),
Pasteur pipettes, isotonic saline, centrifuge, timer, mixing
sticks, eryclone anti-human globulin (Coomb’s) reagent,
Eryclone anti-D (lgG) or RHOFINAL anti-D (lgM + lgG).
336 Concise Book of Medical Laboratory Technology: Methods and Interpretations Test Procedure
Bring reagent and samples to room temperature before
1. Place one drop of Eryclone anti-D (lgM) reagent on a
2. Pipette one equal drop of whole blood on the slide.
3. Mix well with a mixing stick uniformly over an area of
4. Rock the slide gently, back and forth.
5. Observe for agglutination macroscopically at two
1. Prepare a 5% suspension of the red cells to be tested
2. Place one drop of Eryclone anti-D (lgM) reagent into
3. Pipette into the test tube one drop of the 5% cell
4. Centrifuge for 1 minute at 1000 rpm (125 g) or 20
5. Gently resuspend the cell button, observing for
agglutination macroscopically.
1. Prepare a 5% suspension of the red cells to be tested
2. Place one drop of any incomplete anti-D (IgG class)
reagent such as Eryclone anti-D (IgG) into a labeled
3. Add to the test tube one drop of the 5% cell suspension
and mix well. Incubate at 37°C for 15 minutes.
4. Wash the contents of the tube thoroughly, at least three
times, with isotonic saline and decant completely after
5. Add two drops of Eryclone anti-human globulin
6. Centrifuge for 1 minute at 1000 rpm (125 g) or 20
7. Very gently, resuspend the cell button and observe for
agglutination macroscopically.
a. Agglutination is a positive test result and indicates the
presence of D antigen. Do not interpret peripheral
b. Cord cells heavily sensitized with anti-D may give a
false negative immediate spin test result.
a. Agglutination indicates the presence of Du antigen
(Presence of weak/partial Ds). No agglutination indicates
the absence of Du antigen (Absence of weak/partial Ds).
b. Mixed field agglutination in the Du test on red cells
from a recently delivered woman may indicate a
mixture of maternal RhoD negative and fetal RhoD
c. Red cells demonstrating a positive direct antiglobulin
test cannot be accurately tested for Du antigen
(Presence of weak/partial Ds).
As under centrifugation or overcentrifugation could
lead to erroneous results, it is recommended that each
laboratory calibrate its own equipment and determine
the time required for achieving the desired results. It is
strongly recommended that as a routine quality control
measures known RhoD positive and RhoD negative red
cells be occasionally run, preferably on a daily basis so as
No comments:
Post a Comment
اكتب تعليق حول الموضوع