Critical requirement of MNPT in the derivation of INR
ISI value of PT used and method of clot detection:
methods used for clot detection ideally close to 1.0 as possible
Accuracy of the INR system depends on:
• Derivation of the correct MNPT and use of mean normal range in each laboratory
• Usage of thromboplastin reagents with ISI of preferably 1.0 or as close to 1.0 as possible
• The correct use of formula to compute the INR
• Uniform understanding of the INR system by clinicians as well as laboratarians
Uniplastin/Liquiplastin/Lyoplastin
Problem: Prolonged Clotting Time
ideal anticoagulant since factor V and factor VII are more stable in citrate. Therefore, citrate
2. Plasma not tested immediately Plasma must be tested within 3 hours of blood collection
the entire vial. Incubate only the requisite quantity
used for the test, may be anticoagulant to blood ratio must be readjusted using the formula,
incorrect C = 1.85 × 10-3(100-H)V
Clinical Hematology: Bleeding Disorders 311
C = Volume of sodium citrate in mL
V = Volume of whole blood + sodium citrate in mL
5. The manufacturer’s test Adhere to the instrument manufacturer’s instructions and test protocol
6. Temperature of water bath not The temperature of the water bath should be preset at 37 ± 1°C
7. Expired reagents are used for testing Check the expiry date of the reagents before use
as per the instructions given in the package insert. Exactly 0.1 mL of plasma and 0.2 mL of
glassware used for testing turbidity. Ensure that clean and dry glassware and micropipettes are used
tips used for retrieving the reagent
on the type and dosage of anticoagulants and also on the time
and last dose of anticoagulant
tation of test results: per the test protocol
stopped as soon as the difference of 2–3 seconds between the beginning and end of clot formation
b. Clot formation is observed There should be adequate light while observing for clot formation
at the base of the tube the test results.
reagent is stored at the recommended temperature
after prolonged storage at 2-8°C Homogenize the reagent by resuspending before use
ments, the entire reagent vial is be incubated
deterioration of the reagent on
1. Broken glassware used allowing Ensure that broken glassware is not used for testing purposes
silica to trigger the clotting reaction
lag between the last dose of the anticoagulant and specimen collection
Problem: Prolonged Clotting Time
ideal anticoagulant since factor V and factor VII are more stable in citrate. Therefore, citrate
2. Plasma not tested immediately Plasma must be tested within 3 hours of blood collection
used for the test, may be the anticoagulant to blood ratio must be readjusted using the formula,
incorrect C = 1.85 × 10–3 (100-H)V.
C = volume of sodium citrate in mL
V = volume of whole blood + sodium citrate in mL
6. Contaminated/wet micropipette Ensure that clean and dry tips are used for retrieving the reagent.
tips used for retrieving the reagent
minimum time period of 3 minutes obtain accurate results
tation of test results. per the test protocol
Clinical Hematology: Bleeding Disorders 313
b. Clot formation is observed There should be adequate light while observing for clot formation
at the base of the tube results
10. Temperature of water bath not The temperature of the water bath should be set at 37 ± 1°C
anticoagulant therapy depends anticoagulant used as well as the time lag between the specimen
on the type and dosage of anti- collection and the last dose of anticoagulant
coagulants and also on the time
lag between specimen collection
and the last dose of anticoagulant
Problem: Shortened Clotting Time
1. Broken glassware allowing Ensure that broken glassware is not used for testing purposes
silica to trigger the clotting
314 Concise Book of Medical Laboratory Technology: Methods and Interpretations FIBRINOGEN ESTIMATION
Problem: Prolonged Clotting Time
ideal anticoagulant since factor V and factor VII are more stable in citrate. Therefore, citrate
2. Plasma not tested immediately Plasma must be tested within 3 hours of blood collection.
plasma and reagents be incubated for at least 10 minutes at 37°C
Do not incubate the entire vial. Incubate only the requisite quantity.
used for the test, may be the anticoagulant to blood ratio must be readjusted using the formula,
incorrect C = 1.85 × 10-3 (100-H) V
C = volume of sodium citrate in mL
V = volume of whole blood-sodium citrate in mL
tation of test results interpreted as per the test protocol
• The stopwatch is not The stopwatch should be stopped as soon as the gel clot is seen since
stopped as soon as the gel there is a difference of 2–3 seconds between the beginning and end of
• Clot formation is observed There should be adequate light while observing for clot formation
• Water droplets are present Wipe the base of the tube to remove additional water as it hampers
at the base of the tube reading of the test results
6. Temperature of water bath not The temperature of the water bath should be set at 37 ± 1°C
No comments:
Post a Comment
اكتب تعليق حول الموضوع