-absorbent canister, the greatest amount of carbon monoxide is produced by

which of the following volatile agents?

A. Sevoflurane

B. Halothane

C. Isoflurane

D. Desflurane

CHAPTER 3 ANSWERS

1. C. Pipeline gases are supplied at pressures between 45 and 55 psi. This is in

contrast to cylinder gas pressures, which are much higher, and are reduced by

pressure regulators to less than 50 psi.

2. D. The fail-safe valve automatically closes nitrous oxide (and other gases) to

prevent delivery of hypoxic gas mixture to the patient. The fail-safe valve is

designed to be activated when oxygen pressure falls below 25 psi.

3. C. The oxygen-flush valve provides gas flow at pipeline pressures of about 45 to

55 psi at 35 to 75 L/min. The high flow of oxygen is provided directly to the common

gas outlet, bypassing the flowmeters and vaporizers. One should be careful when

using the oxygen-flush valve, as high gas flows at high pressures can cause lung

barotrauma in the patient.

4. A. Gas flowmeters are calibrated for a particular gas. Gas flow rate depends on its

viscosity at low laminar flows, and its density at high turbulent flows. Flowmeters

are tapered in shape, with the diameter the smallest near the bottom of the tube.

5. B. The oxygen flowmeter is situated nearest to the gas outlet. This is because, if a

leak develops in the flowmeter tubes, a hypoxic gas mixture can be delivered to the

patient. To minimize this, the oxygen flowmeter is positioned downstream and

nearest to the gas outlet.

6. D. Modern vaporizers are agent-specific and temperature-compensated. Also,

specific fillers are available for each volatile agent, which prevent filling on the

wrong agent. A constant concentration of agent is delivered, unaffected by

temperature or flow rates. Temperature compensation is achieved by a metallic strip

composed of two different metals, which expands/contracts to deliver a constant

concentration of vapor.

7. A. The Tec 6 desflurane vaporizer is electrically heated to 39°C and pressurized to

2 atm. This is done because desflurane boils at room temperature at sea level (1

atm). The heating and pressurization optimizes the delivery of desflurane.

8. B. Vaporizers are located between the flowmeters (upstream) and the common gas

outlet (downstream). In other words, vaporizers are located outside the circle system.

This decreases the likelihood of delivery of high vapor concentrations when using the

oxygen-flush valve.

9. A. An ascending bellow collapses when disconnection occurs. A descending

bellow, however, continues to fill by gravity when disconnection occurs. Therefore,

ascending bellows are preferred for anesthesia ventilators.

10. B. NIOSH recommends limiting operating-room concentration of nitrous oxide to

25 ppm. Minimizing operating-room pollution is important to prevent health-related

effects in health-care providers. Waste-scavenging systems are utilized to decrease

operating-room pollution.

11. D. NIOSH recommends limiting operating-room concentration of volatile agents to

2 ppm. Minimizing operating-room pollution is important to prevent health-related

effects in health-care providers. Waste-scavenging systems are utilized to decrease

operating-room pollution.

12. C. The capacity of an “E” cylinder of oxygen is about 625 to 700 L. The pressure

in a full cylinder is about 1,800 psi at 20°C. Cylinders are color-coded, with oxygen

being green, nitrous oxide being blue, and air being yellow.

13. D. Pressure in a half-full “E” cylinder of nitrous oxide will still be 745 psi. Nitrous

oxide is present in the cylinder as a liquid, and therefore, the volume remaining in the

cylinder does not reflect the pressure in the cylinder. Capacity of an “E cylinder” of

nitrous is about 1590 L. It is not until three-fourth of the gas is consumed (about 400

L remaining) that the pressure in the cylinder begins to fall. Therefore, the reliable

way to determine the remaining nitrous oxide in the cylinder is to weigh the cylinder.

The empty weight of the cylinder is stamped on the cylinder.

14. B. Cylinder manufactures have adopted the pin index safety system, which

prevents attachment of wrong gas cylinder to the anesthesia machine. The diameter

index safety system prevents attachment of the wrong gas hose from the wall supply.

Hanger yoke assembly is the method of attachment of gas cylinders to the anesthesia

machine.

15. B. A line-isolation monitor, when alarming, indicates that a single fault has

occurred between the power line and the ground. As soon as the alarm is triggered,

the equipment should be checked, especially the last equipment that was plugged in.

A single fault does not cause an electrical shock, as two faults are required to

produce a shock.

16. A. The highest content of soda lime is calcium hydroxide (75%). Other constituents

include sodium (3%) and potassium hydroxide (1%), water (20%), and silica, which

is added to produce hardness. An indicator dye, such as ethyl violet, is added to

indicate the degree of exhaustion.

17. D. End products of the reaction occurring in a soda lime CO2 canister are

carbonates, sodium hydroxide (regeneration), water, and heat. Following are the

reactions:

18. B. Advantages of a circle system include the use of low fresh gas flow rates

because of the presence of a CO2

-absorbent canister. However, if the CO2 absorbent

is exhausted during a surgical procedure, the fresh gas flow rate has to be increased.

A minimum fresh gas flow rate of 5 L/min will make the use of the absorbent

unnecessary. Newer anesthesia machines allow changing the CO2

-absorbent canister

during the surgical procedure, if necessary.

19. D. Advantages of the circle system include economy (low fresh gas flow rates,

decreased use of volatile agents), conservation of heat and humidity, and decreased

operating-room pollution. Disadvantages of circle system include greater size,

decreased portability, increased risk of disconnection, and increased resistance to

patient breathing.

20. A. While resuscitation devices such as Ambu bags or bag-mask units have

nonrebreathing valves, neither the Mapleson (only has adjustable pressure-limiting

valve) nor the circle system (only has unidirectional valves and does allow

rebreathing) has this component. Ambu resuscitation bags do allow for positivepressure ventilation as the intake valve closes during bag compression. The patient

valve has low resistance, but can become obstructed by exhaled moisture. Ambu

bags have a reservoir system to prevent room air entrapment and are able to deliver

nearly 100% oxygen.

21. D. One advantage of the circle system ventilation when compared to the Mapleson

system is the presence of unidirectional valves (inspiratory and expiratory valves).

With the use of such valves, the volume of dead-space ventilation is limited only to

that volume distal to Y-piece (including the endotracheal tube), where inspiratory and

expiratory gases mix and converge, regardless of the length of tubing proximal to the

Y-piece (to the anesthesia machine).

22. C. Malfunction in either of the unidirectional valves within a circle system could

result in the accumulation and eventual CO2

rebreathing that may result in

hypercapnia.

23. A. During spontaneous ventilation/breathing of the patient, the Mapleson circuit

providing for the most efficacy ranges from A > D > C > B (in the order of

decreasing efficiency).

24. A. The efficiency of Mapleson systems drops from D > B > C > A for controlled

ventilation. The Mapleson D circuit is most efficient during controlled ventilation, as

its fresh gas flow drives expired air away from the patient and toward the

expiratory/exhaust valve.

25. A. Sevoflurane is degraded by soda lime, resulting in the production of a

potentially nephrotoxic compound A. Compound A production is increased by using

low fresh gas flow rates, using high concentrations of sevoflurane, and for long hours

(>6 hours).

26. D. Desflurane produces the highest amount of carbon monoxide in the CO2

-

absorbent canister, which can increase carboxyhemoglobin blood concentration.

Production of carbon monoxide is increased by using low fresh gas flow rates, high

concentrations of volatile agent, and a dry absorbent.

Patient Monitoring

Darren Hyatt, Ala Nozari, and Edward Bittner

1. To help encourage universal quality and safety practices, the ASA has adopted and

mandates the use of all the following monitors during general anesthesia, except

A. An oxygen analyzer

B. Capnography

C. Continuous visual display of an ECG

D. A peripheral nerve stimulator

2. Current ASA standards require that during anesthesia, systemic blood pressure and

heart rate be evaluated at least every

A. 3 minutes

B. 5 minutes

C. 7 minutes

D. 10 minutes

3. Lead II of an ECG is represented by placing the

A. Positive electrode on the right arm and the negative electrode on the left leg

B. Negative electrode on the right arm and the positive electrode on the left leg

C. Positive electrode on the right arm and the negative electrode on the left arm

D. Negative electrode on the right arm and the positive electrode on the left arm

4. During the course of a complicated cardiac case, the surgeon informs you that he is

worried about damage to the right coronary artery in a patient with a right-dominant

coronary system. During reperfusion, you are looking for signs of ischemia, and are

most interested in leads

A. V1–V3

B. V4–V6

C. II, III, and AvF

D. I and AvL

5. Use of lead V5 alone on ECG results in the detection of _____ (%) of ischemic

episodes:

A. 35

B. 55

C. 75

D. 95

6. You are taking over a case from another anesthesia provider with a patient in the

beach chair position and a history of moderate carotid artery disease. You are told

during pass-off that the patient’s blood pressures have consistently been 90/50 mm

Hg. You notice the blood pressure cuff on the left arm is one or two sizes small and

barely stays on the patient. A blood pressure cuff that is too small will

A. Incorrectly underestimate the true blood pressure

B. Incorrectly overestimate the true blood pressure

C. Randomly both over- and underestimate the true blood pressure

D. Not give an incorrect blood pressure, but will be uncomfortable in an awake

patient

7. When performing the oscillometric method to measure blood pressure, for example,

when you do not have a stethoscope or automated blood pressure cuff, it is important

to remember that you will not be able to measure the

A. Systolic blood pressure

B. Diastolic blood pressure

C. Mean arterial pressure

D. Diastolic or mean arterial blood pressure

8. The diastolic blood pressure recorded with an automated blood pressure cuff using

the oscillometric method will be

A. Approximately 10 mm Hg higher when compared to direct arterial measurement

B. Approximately 10 mm Hg lower when compared to direct arterial measurement

C. Equal to direct arterial measurement

D. Random and unreliable

9. When measuring blood pressure manually and listening for Korotkoff sounds, the

diastolic blood pressure is measured at the onset of

A. Phase 1

B. Phase 2

C. Phase 3

D. Phase 5

10. You are preparing for an emergent mitral valve repair that will need to be done on

cardiopulmonary bypass (CPB). While on CPB

A. A pulse oximeter can be used to monitor oxygen saturation

B. A noninvasive blood pressure cuff can be used to monitor perfusion pressures

C. An arterial line can be used to measure perfusion pressures

D. None of the above

11. The incidence of distal ischemia resulting from arterial cannulation is less than

A. 10%

B. 1%

C. 0.1%

D. 0.01%

12. When considering the advantages and disadvantages of different sites for arterial

cannulation such as radial, ulnar, femoral, brachial, and dorsalis pedis, the

A. Radial artery provides the principal source of blood to the hand

B. Cannulation of ulnar artery is commonly associated with damage to the median

nerve

C. Dorsalis pedis artery is commonly used during emergencies and low-flow states

D. Cannulation of the femoral artery risks local and retroperitoneal hematoma

13. Systolic blood pressures are generally higher and diastolic blood pressures are

generally lower in which of the following conditions?

A. The further you are from the heart when using a direct arterial measurement

B. The closer you are to the heart when using a direct arterial measurement

C. When using an automated noninvasive blood pressure cuff compared to a direct

arterial measurement

D. When recording from an over dampened arterial tracing

14. While taking care of a patient, you notice that the arterial monitor transducer has

slipped off its stand and is hanging approximately 30 cm lower than where it was

originally leveled. This would correspond to a blood pressure reading that is

A. 30 mm Hg lower than the actual pressure

B. 30 mm Hg higher than the actual pressure

C. 22 mm Hg lower than the actual pressure

D. 22 mm Hg higher than the actual pressure

15. An important consideration in using the subclavian approach for central venous

access includes the

A. Ease of compressibility if a hematoma or laceration develops

B. Lower risk of pneumothorax when compared to internal jugular approach

C. Ability of the vessel to remain patent in the setting of hypovolemia

D. Increased risk of damaging the brachial plexus when compared to internal

jugular approach

16. When interpreting a CVP waveform, the end of systole best coincides with the

A. A wave

B. C wave

C. V wave

D. X decent

17. When interpreting a CVP waveform, the beginning of systole is best represented by

the

A. A wave

B. C wave

C. V wave

D. X decent

18. After placing a central line in an unstable patient in the ICU, you notice the initial

CVP tracing shows very prominent C–V waves. If an echocardiogram was then

obtained, you might expect to find

A. Cardiac tamponade

B. Significant tricuspid regurgitation

C. Atrial fibrillation

D. AV dissociation

19. You receive a patient from the emergency department with multiple stab wounds to

the upper abdomen. The patient is unstable, and needs to emergently come to the

operating room with minimal to no time for fluid resuscitation. After placing a central

line, you notice loss of the Y descent on the CVP tracing, as well as universally

elevated filling pressures. If you were to then do an echocardiogram, you might

expect to find which of the following?

A. Cardiac tamponade

B. Significant tricuspid regurgitation

C. Descending thoracic aortic dissection

D. AV dissociation

20. The risk of complication from pulmonary artery catheter placement is less than

A. 0.05%

B. 0.5%

C. 5%

D. 15%

21. Insertion of a pulmonary artery catheter can be beneficial in the management of all of

the following cases, except

A. Helping to determine cardiogenic versus noncardiogenic pulmonary edema

B. Following cardiac output in an unstable patient with acute-onset tricuspid

regurgitation

C. Following the response to therapy in a patient with severe pulmonary

hypertension

D. Following response to therapy in an unstable septic patient using mixed venous

oxygen tension

22. During placement of a pulmonary artery catheter, you are watching the pressure

tracing, as shown. At the point indicated by the arrow, the catheter tip is located in

the

Figure 4-1.

A. Right atrium

B. Right ventricle

C. Pulmonary artery

D. Wedge position

23. The tip of a pulmonary artery catheter typically enters the pulmonary artery at

approximately

A. 15 to 25 cm

B. 25 to 35 cm

C. 35 to 45 cm

D. 45 to 55 cm

24. Typical mixed venous oxygen tension in a healthy adult is

A. 25 mm Hg

B. 40 mm Hg

C. 55 mm Hg

D. 75 mm Hg

25. A pulmonary artery catheter is placed to help guide management of hypotension.

Cardiac output is found to be markedly decreased with low central venous,

pulmonary artery, and pulmonary artery occlusion pressures. Systemic vascular

resistance is moderately elevated. Of the options listed below, the most beneficial

intervention at this time would be to

A. Administer volume

B. Begin diuresis

C. Start an infusion of milrinone

D. Start an infusion of epinephrine

26. A pulmonary artery catheter is placed to help guide management of an obese patient

with a known history of poorly controlled obstructive sleep apnea who is admitted

with refractory hypotension. Cardiac output and pulmonary artery occlusion

pressures are markedly decreased, while central venous and pulmonary artery

pressures are markedly increased. Of the options listed below, the most beneficial

intervention at this time would be to

A. Administer volume

B. Begin diuresis

C. Start an infusion of milrinone

D. Start an infusion of epinephrine

27. Normal systemic vascular resistance ranges between ______ (dynes)(s)/cm5

:

A. 50 and 150

B. 300 and 600

C. 900 and 1500

D. 1800 and 2100

28. Normal pulmonary vascular resistance ranges between ______ (dynes)(s)/cm5

:

A. 50 and 150

B. 300 and 600

C. 900 and 1500

D. 1800 and 2100

29. The cardiac index in a healthy adult ranges between ______ L/min/m2

:

A. 0.8 and 1.2

B. 1.4 and 2.0

C. 2.2 and 4.2

D. 4.4 and 6.0

30. Serious complications with transesophageal echocardiography (TEE), such as oral or

pharyngeal injury or esophageal rupture, have an incidence as high as

A. 0.01%

B. 0.1%

C. 1%

D. 10%

31. When evaluating regurgitant lesions with transesophageal echocardiography, the

Nyquist limit should be set between ______ cm/s:

A. 30 and 40

B. 40 and 50

C. 50 and 60

D. 60 and 70

32. When evaluating flow at a specific point during echocardiography, you would use

A. Continuous-wave Doppler

B. Pulse-wave Doppler

C. Color Doppler

D. Pulse-wave or continuous-wave Doppler

33. Pulse oximetry illuminates tissue samples with two wavelengths of light in order to

calculate oxygen saturation. These wavelengths are ______ nm:

A. 540 and 780

B. 660 and 940

C. 720 and 960

D. 480 and 720

34. The accuracy of pulse oximetry can be significantly reduced by all of the following,

except

A. Intravenous bolus of methylene blue

B. Intravenous bolus of heparin

C. Severe acidosis

D. Low blood flow

35. A patient with carboxyhemoglobin will have a pulse oximetry reading that

A. Converges around a saturation of 85%

B. Converges around a saturation of 65%

C. Converges around a saturation of 45%

D. Varies widely

36. A patient with methemoglobinemia will have a pulse oximetry reading that

A. Converges around a saturation of 85%

B. Converges around a saturation of 65%

C. Converges around a saturation of 45%

D. Varies widely

37. For the removal of a complex spinal cord tumor, the surgeon expresses concern of

damage to the anterior spinal artery. The monitoring that would be helpful to

determine viability of the anterior spinal cord intraoperatively would include

A. Electroencephalography

B. Motor-evoked potentials

C. Somatosensory-evoked potentials

D. Bispectral index or Sedline monitoring

38. A sudden drop in somatosensory-evoked potentials (SSEPs) would cause you to be

worried about

A. Damage to the anterior spinal artery

B. Damage to the posterior spinal arteries

C. An insufficient depth of anesthesia

D. The inadvertent administration of a neuromuscular blocking agent

39. During cervical spine surgery for the resection of an intradural mass, the patient

begins to cough. The concentration of isoflurane is subsequently increased. With

respect to somatosensory-evoked potential (SSEP) monitoring, you would expect

A. Amplitude and latency to decrease

B. Amplitude and latency to increase

C. Amplitude to decrease and latency to increase

D. Amplitude to increase and latency to decrease

40. While monitoring somatosensory-evoked potentials, an increase in amplitude is

noted. Of the options listed below, the most likely medication to have caused this

increase in amplitude would be

A. Etomidate

B. Propofol

C. Midazolam

D. Sevoflurane

41. If somatosensory-evoked potentials change significantly, the anesthesia provider

should consider

A. Increasing blood pressure

B. Hyperventilating the patient

C. Cooling the patient

D. Hemodilution

42. In the capnogram below, the segment that correlates with the exhalation of anatomic

dead space is represented by points

Figure 4-2.

A. A to B

B. A to C

C. C to D

D. D to E

43. In the capnogram (Fig. 4-2), the segment correlating with inspiration is represented

by points

A. A to B

B. A to C

C. C to D

D. D to E

44. Capnography can help detect all of the following, except

A. Endobronchial intubation

B. Esophageal intubation

C. Bronchospasm

D. Pulmonary embolism

45. The capnograph depicted in Figure 4-3 is most likely a result of

Figure 4-3.

A. Pulmonary embolism

B. Bronchospasm or airway obstruction

C. Esophageal intubation

D. Elimination of neuromuscular blockers

46. Approximately 30 minutes after the induction of general anesthesia in a healthy adult

patient, you notice that core body temperature has dropped by a full degree Celsius.

This is most likely due to

A. Conduction

B. Convection

C. Redistribution

D. Radiation

47. According to the American Society of Anesthesiologists, temperature monitoring is

A. Always required

B. Never required, but recommended

C. Required for all general anesthetics, however not required for sedation

D. Up to the discretion of the anesthesia provider

48. Detrimental effects of hypothermia include all of the following, except

A. Increasing cerebral oxygen consumption

B. Increasing surgical site infections

C. Impairment of platelet function

D. Increasing the duration of action of muscle relaxants

49. During a complex mitral valve replacement, it is determined that the patient will

benefit from brief protective hypothermia. Of the options listed below, core

temperature is best measured via the

A. Tympanic membrane

B. Bladder

C. Nasopharnyx

D. Rectum

50. While monitoring a patient for return of neuromuscular function after using

rocuronium, you notice the patient has regained four twitches using train of four

stimulations. With four twitches on train of four stimulations, the patient may still

have blockage of acetylcholine receptors of up to

A. 25%

B. 50%

C. 75%

D. 90%

CHAPTER 4 ANSWERS

1. D. ASA standards mandate the use of pulse oximetry, capnography, an oxygen

analyzer in the breathing system, disconnect alarms, a visual display of an ECG,

systemic blood pressure and heart rate monitoring, and temperature monitoring (when

clinically indicated) for all cases. The use of a peripheral nerve stimulator is not a

mandated monitor.

2. B. During the delivery of anesthesia, the current standard of care is to measure

systemic blood pressure and heart rate every 5 minutes at a minimum. The clinical

scenario and phase of the operation may mandate more frequent monitoring, which is

up to the judgment of the anesthesia provider.

3. B. Lead I correlates with the placement of the negative electrode on the right arm

and the positive electrode on the left arm. Lead II correlates with the placement of the

negative electrode on the right arm and the positive electrode on the left leg. Lead III

correlates with the placement of the negative electrode on the left arm and the

positive electrode on the left leg.

4. C. The understanding of coronary anatomy and regions of ischemia on an ECG is

fundamental. The right coronary artery provides perfusion to the inferior of the heart

in approximately 80% of patients who are considered to be right-dominant (the

posterior descending artery is supplied by the right coronary artery in a rightdominant system). This inferior distribution is represented by leads II, III, and AvF.

The anterior wall is supplied by the left anterior descending artery, and is represented

roughly by leads V1–V4. The lateral wall of the heart is supplied primarily by the

left circumflex artery, and is represented by I, AvL, V5, and V6.

5. C. The use of the V5 lead results in the detection of 75% of ischemic episodes.

This can be increased to 90% with the addition of the V4 lead, and up to 96% with

the addition of leads II and V4.

6. B. A properly-sized noninvasive blood pressure cuff should encompass 40% of the

circumference of the arm. A cuff that is too small will result in a reading that is

incorrectly high, whereas a cuff that is too large will result in a lower-than-accurate

pressure. This is particularly worrisome in this patient when considering her cerebral

perfusion pressure, since she already has a history of carotid artery disease and is in

the beach chair position.

7. B. When using the oscillometric method to measure blood pressure, the cuff is

inflated until no oscillations on the sphygmomanometer are seen. The cuff is then

slowly deflated until oscillations are seen, which represents the systolic blood

pressure. As the cuff continues to be deflated, you note the point where maximal

oscillations occur. This point of maximal oscillation represents the mean arterial

pressure. It is not possible to measure a diastolic blood pressure with the

oscillometric method.

8. A. The DINAMAP (device for indirect noninvasive automatic mean arterial

pressure) method for measuring blood pressure uses an automated cuff that measures

oscillometric variations with reduction in cuff pressure to calculate systolic, mean,

and diastolic pressures. In general, diastolic measurements with DINAMAP are

about 10 mm Hg higher with automated as opposed to direct arterial measurement,

whereas systolic and mean pressures tend to correlate well.

9. D. Korotkoff sounds are used to interpret blood pressure when using a stethoscope

and a noninvasive blood pressure cuff, and is described in 5 phases of sound. Phase

1 heralds the onset of the first sound heard and correlates with the systolic blood

pressure. Phase 5 occurs at the cuff pressure at which the sound first disappears, and

is the phase recommended by the American Heart Association to correspond most

reliably with the diastolic heart sound. In cases where Phase 5 does not occur (the

sound never fully disappears), Phase 4 is then used to represent the diastolic blood

pressure, and is described as a thumping or muting of the sound just before diastole.

Phases 2 and 3 have no clinical significance.

10. C. Both pulse oximetry and noninvasive blood pressure cuffs require pulsatile

blood flow in order to obtain measurements. These monitors will not be effective

during CPB when blood flow is artificially sustained with a more continuous flow.

This can also be the case with some patients on left ventricular assist devices, and

venous to arterial extracorporeal membrane oxygenation devices, where pulsatile

flow is minimal.

11. C. Complications from arterial cannulation include distal ischemia (<0.1%),

infection, and hemorrhage. Common sites for cannulation include radial, brachial,

axillary, dorsalis pedis, and femoral arteries. Common indications for direct blood

pressure monitoring include cardiopulmonary bypass, when wide swings in BP are

expected, when rigorous control of BP is necessary, and when there is need for

multiple arterial blood gas measurements.

12. D. The ulnar artery is the principal source of blood flow to the hand. Hence radial

artery cannulation is much more commonly used for invasive blood pressure

monitoring. Cannulation of the brachial artery risks damage to the median nerve. The

femoral artery is often used in emergencies, since it is a large vessel and can still be

identified in low flow states. Cannulation of the femoral artery risks both local and

retroperitoneal hematoma. Dorsalis pedis artery cannulation, while not ideal since it

is far from the central circulation, can reliably measure mean arterial pressure.

13. A. Systolic blood pressures are generally higher and diastolic blood pressures are

generally lower the further you are from the heart when using direct invasive arterial

measurement. For example, when comparing a dorsalis pedis arterial measurement to

a femoral arterial measurement, the dorsalis pedis will record higher systolic and

lower diastolic pressures compared to the femoral line. However, the mean arterial

pressures will be approximately the same. A noninvasive automated blood pressure

cuff will tend to correlate with systolic arterial blood pressures, but the diastolic

pressure will be approximately 10 mm Hg lower when measured via the direct

invasive arterial monitor. An over dampened arterial line tracing will tend to reduce

systolic pressures and increase diastolic pressures.

14. D. For every 30 cm in height that a transducer is moved up and down, there is a

corresponding change of 22 mm Hg in the blood pressure reading (1 cm H2O = 0.74

mm Hg).

15. C. Risks and benefits of different central cannulation sites are important for an

anesthesia provider to understand. The internal jugular approach has good landmarks,

predictable anatomy, and the convenience of being easily accessible at the head of

the bed. Disadvantages include risk of carotid artery puncture, trauma to the brachial

plexus, and risk of pneumothorax with lower placements. The left internal jugular

vein carries the added risk of damage to the thoracic duct, and can be more difficult

to pass a pulmonary artery catheter when needed. The external jugular vein can also

be cannulated, and its superficial location makes it an easy target, but it can be more

difficult to thread a catheter centrally. The subclavian approach has the benefit of

also having good landmarks, as well as remaining relatively patent in a hypovolemic

patient. The subclavian however does carry the highest risk of pneumothorax, and

can be difficult to compress if a hematoma or laceration occurs.

16. C. In the CVP waveform depicted below, the A wave represents atrial contraction,

the C wave represents bulging of the tricuspid valve into the atrium during the

beginning of systole, the X decent occurs during systole and corresponds to atrial

relaxation, the V wave represents filling of the atrium while the tricuspid valve is

closed, and the Y descent occurs when the tricuspid valve opens and the atrium starts

to empty.

Figure 4-4.

17. B. In the CVP waveform depicted in Figure 4-4, the A wave represents atrial

contraction, the C wave represents bulging of the tricuspid valve into the atrium

during the beginning of systole, the X decent occurs during systole and corresponds

to atrial relaxation, the V wave represents filling of the atrium while the tricuspid

valve is closed, and the Y descent occurs when the tricuspid valve opens and the

atrium starts to empty.

18. B. During systole in a patient with tricuspid regurgitation, part of the ejected

volume flows backward into the atrium. Instead of seeing a small C wave that

normally represents the bulging of the tricuspid valve, a much larger C wave would

be seen as blood flows retrograde into the right atrium and toward the transducer.

This retrograde blood flow would continue throughout the systole, and would,

therefore, also increase the V wave size, since this is a systolic component of the

CVP trace. During cardiac tamponade, there will be elevated pressures throughout

the entire waveform, as well as loss of the Y descent. In patients with atrial

fibrillation, there will be a loss of the A wave, since there is no longer a uniform

atrial contraction, and an overall increase in the C wave size, since filling pressures

elevate to compensate and improve ventricular filling. With AV dissociation, there

are large and exaggerated A waves (often called “cannon” A waves), which represent

atrial contraction against a closed tricuspid valve.

19. A. See the answer explanation of Question 18. It would be highly unlikely to have

elevated filling pressures in a bleeding trauma patient who has not yet been

resuscitated. Aortic dissections can cause cardiac tamponade, but only if they

involve the aortic root and then extend into the pericardium.

20. B. While the incidence of complications is infrequent, some of the complications

can carry severe morbidity and mortality risks. In addition to universal complications

associated with central line placement, some additional pulmonary artery catheter

complications include dysrhythmias (most common), catheter knotting, cardiac valve

injury, pulmonary artery rupture, development of complete heart block in a patient

with preexisting left bundle branch block, pulmonary thromboembolism or air

embolism, bacteremia, endocarditis, and sepsis.

21. B. As well as knowing some valuable indications, it is important to know some of

the limitations of a pulmonary artery catheter before subjecting a patient to risks. For

example, the measurement of cardiac output in patients with tricuspid regurgitation or

ventricular septal defects is inaccurate due to dilution of the injectate. Pulmonary

artery occlusion pressure can also inaccurately represent left ventricular end

diastolic pressure in patients with mitral stenosis, left atrial myxomas, pulmonary

venous obstruction, elevated alveolar pressures, and decreased left ventricular

compliance. Other common errors in measurement that are not patient dependent can

include an inaccurate volume or temperature of the injectate solution.

22. B. A pulmonary artery catheter is placed while monitoring the pressure changes

measured at the tip of the catheter. The first section shows a traditional CVP

waveform measured in the right atrium. As the catheter is advanced, a systolic stepup is seen when entering the right ventricle, a diastolic step-up when entering the

pulmonary artery, and a return to a traditional CVP waveform when entering the

wedge position.

Figure 4-5.

Comments

Search This Blog

Archive

Show more

Popular posts from this blog

TRIPASS XR تري باس

CELEPHI 200 MG, Gélule

ZENOXIA 15 MG, Comprimé

VOXCIB 200 MG, Gélule

Kana Brax Laberax

فومي كايند

بعض الادويه نجد رموز عليها مثل IR ، MR, XR, CR, SR , DS ماذا تعني هذه الرموز

NIFLURIL 700 MG, Suppositoire adulte

Antifongiques مضادات الفطريات

Popular posts from this blog

علاقة البيبي بالفراولة بالالفا فيتو بروتين

التغيرات الخمس التي تحدث للجسم عند المشي

إحصائيات سنة 2020 | تعداد سكَان دول إفريقيا تنازليا :

ما هو الليمونير للأسنان ؟

ACUPAN 20 MG, Solution injectable

CELEPHI 200 MG, Gélule

الام الظهر

VOXCIB 200 MG, Gélule

ميبستان

Popular posts from this blog

TRIPASS XR تري باس

CELEPHI 200 MG, Gélule

Popular posts from this blog

TRIPASS XR تري باس

CELEPHI 200 MG, Gélule

ZENOXIA 15 MG, Comprimé

VOXCIB 200 MG, Gélule

Kana Brax Laberax

فومي كايند

بعض الادويه نجد رموز عليها مثل IR ، MR, XR, CR, SR , DS ماذا تعني هذه الرموز

NIFLURIL 700 MG, Suppositoire adulte

Antifongiques مضادات الفطريات

Popular posts from this blog

Kana Brax Laberax

TRIPASS XR تري باس

PARANTAL 100 MG, Suppositoire بارانتال 100 مجم تحاميل

الكبد الدهني Fatty Liver

الم اسفل الظهر (الحاد) الذي يظهر بشكل مفاجئ bal-agrisi

SEDALGIC 37.5 MG / 325 MG, Comprimé pelliculé [P] سيدالجيك 37.5 مجم / 325 مجم ، قرص مغلف [P]

نمـو الدمـاغ والتطـور العقـلي لـدى الطفـل

CELEPHI 200 MG, Gélule

أخطر أنواع المخدرات فى العالم و الشرق الاوسط

Archive

Show more