Chapter 8 ■ Cardiorespiratory Monitoring 53
(1) Typically utilizes the same electrodes as are
(2) Signal path usually from right arm (white) to left
arm (black) electrodes, although some monitors
may use right arm (white) to left leg (red or
b. Impedance to the high-frequency signal is measured.
(1) Impedance is the electrical resistance to the signal.
(2) Changes in lung inflation cause an alteration in
the density of the chest cavity, which is detected
(3) Changes in impedance modulate a proportional
change in the amplitude of the high-frequency
c. The change in impedance, as seen by the modulation of the high-frequency signal, is detected and
quantified by the monitor and recorded as breaths
d. The monitor has an impedance threshold limit below
which changes in impedance are not counted as valid
respiratory activity—cardiac pumping with associated
changes in pulmonary blood flow will also cause
Fig. 8.6. Normal P-, QRS-, and T-wave detection. Top: Lead II
tracing with electrodes properly placed. Note normal P-, QRS-, and
T-wave detection. Middle: Lead II tracing with electrodes close
together on anterior chest wall. Note altered QRS- and decreased
T-wave amplitude. Bottom: Lead II tracing with electrodes placed
lateral on the abdomen. Note decreased wave amplitude and flattened P wave.
Fig. 8.7. Transthoracic impedance pneumography: Diagrammatic
representation of the path of the high-frequency signal between
chest wall electrodes. Most monitors transmit signal right arm
(white) → left arm (black), less commonly right arm (white) → left
changes in thoracic impedance (usually much smaller
changes than those associated with respiration).
1. Equipment is the same as that for cardiac monitoring;
multiparameter monitors incorporate both cardiac and
respiratory monitoring into single units.
2. Respiratory monitoring parameters
b. Coincidence alarm with rejection applies when
respiratory rate being detected is equal to the heart
rate activity being detected by the cardiac portion of
c. Default limits should be tailored to the neonatal
54 Section II ■ Physiologic Monitoring
(1) Adjustable apnea time-delay setting (length of
apnea in seconds before alarming)
(2) Typical apnea time delay is 15 to 20 seconds.
1. Include previously discussed precautions for cardiac
2. Muscular activity may be interpreted as respiration,
resulting in failure to alarm during an apneic episode
1. Same as for cardiac monitor
2. Ensure that the respiratory waveform correlates to the
true initiation of inspiration.
4. Set desired low and high respiratory rate and apnea
1. Skin lesions (see H1 under Cardiac Monitoring)
a. Hardware or software failure
a. False-positive “respiratory” signal in the absence of
(1) Chest wall movement with airway obstruction
(2) Nonrespiratory muscular action (i.e., stretching,
seizure, or hiccups) producing motion artifact
b. False apnea alarm despite normal respiratory activity
(1) Improper sensitivity not detecting present respiratory activity
(2) Incorrect electrode placement
c. Inappropriate alarm parameters for patient
4. Accurate assessment of respiratory rate not practical
when using high-frequency ventilatory modes
1. Graphical representation of heart rate and respiratory
of apnea, periodic breathing, and bradycardia
2. Identification of chronologic relationships between
1. Heart rate is plotted graphically as beats per minute
(y-axis) versus time (x-axis).
2. Respiratory waveform is compressed to allow display of
3. Short-term trending allows constant updating as the
oldest information is displaced (typically based on a
4. Time relationship between heart rate and respiratory
a. Allows for visualization of entire apneic episodes
and identification of precipitating factors (e.g., a
drop in respiratory rate may precede bradycardia)
same fashion as the heart rate on a second y axis)
artifact caused by patient’s moving arm coming in contact with
chest electrodes (note change in respiratory frequency signal).
No comments:
Post a Comment
اكتب تعليق حول الموضوع