2. The sphygmomanometer uses a pneumatic cuff to
encircle the upper arm or leg and a pressure gauge
(manometer) to register the pressure in the cuff.
3. There are two types of manometers
b. Aneroid (mechanical air gauge)
4. The encircling pneumatic cuff is inflated to a pressure
higher than the estimated systolic pressure in the
underlying artery. The cuff pressure compresses the
5. A stethoscope placed distal to the cuff, over the
occluded artery, will pick up the Korotkoff sounds as
the cuff is deflated and the pressure of the cuff decreases
to the point at which blood flow resumes through the
6. Korotkoff sounds are the noise generated by blood flow
returning to the compressed artery and originate from a
combination of turbulent blood flow and oscillations of
the arterial wall. The sounds have been classified into
a. Phase I: Appearance of clear tapping sounds corresponding to the appearance of a palpable pulse
b. Phase II: Sounds become softer and longer
c. Phase III: Sounds become crisper and louder
d. Phase IV: Sounds become muffled and softer
e. Phase V: Sounds disappear completely. The fifth
phase is thus recorded as the last audible sound.
7. An 8- to 9-MHz Doppler device can be used in place of
a stethoscope. This device will detect only systolic BP
2. When only intermittent BP measurements are required
1. Severe edema in the limb to be measured will muffle
2. Decreased perfusion, ischemia, infiltrate or injury in
3. Peripheral venous/arterial catheter in limb
1. Provides only intermittent BP measurements
2. Manual measurement cumbersome or impossible in
3. Accuracy depends on ability to recognize Korotkoff
sounds and may be user-dependent.
4. Pressure may not be detectable in low-perfusion states
or shock. Do not assume that it is simply an equipment
problem; use clinical correlation.
5. Pressure is not detectable or is inaccurate when the
baby is actively moving or agitated.
6. Measures only systolic and diastolic BP; mean BP measurement not available
Chapter 9 ■ Blood Pressure Monitoring 57
7. Can be used only to measure pressure in the upper arm
8. The Korotkoff sound method tends to give values for
9. Inaccurate measurements (Table 9.1)
1. Neonatal cuff (Table 9.2). Select a cuff that will fit
without overlapping. The width should be 90% of the
limb circumference at the midpoint (5).
2. Mercury manometer or aneroid-type gauge
3. Appropriate-sized stethoscope with diaphragm or
a. Cuff too small: BP will be higher than actual BP.
b. Cuff too large: BP will be lower than actual BP.
2. Check functional integrity of manometer
3. Check integrity of cuff for leaks
4. Check speed of cuff deflation: If deflation is too rapid,
5. Patient must be quiet and still during measurements.
6. For optimal infection control, use disposable cuff for
1. Place the infant supine, with the limb fully extended
2. Measure the limb circumference and select the appropriate size cuff for the limb.
a. Neonatal cuffs are marked with the size range
b. When the cuff is wrapped around the limb, the
end of the cuff should line up with the range mark
c. If the end of the cuff falls short of the range mark,
d. If the end of the cuff falls beyond the range mark,
3. Apply the cuff snugly to the bare limb, above the elbow
Table 9.1 Sources of Error in Indirect Blood Pressure Measurements
Problem Effect on Blood Pressure Precaution
3. Dry, degraded or cracked tubing
Falsely low values 1. Check level of mercury at zero cuff pressure
2. Check for cleared definition of meniscus
3. Verify that pressure holds when tightened. Check
Inappropriate cuff size Verify appropriately sized cuff
Cuff applied loosely Falsely high values owing to ballooning of bag and
narrowing of effective surface
Cuff applied too tightly Inaccurate reading owing to impedance of flow
Apply cuff snugly without undue pressure
Rapid deflation of cuff 1. Falsely low values owing to inaccurate detection
2. Falsely high values owing to inadequate equilibration
between cuff pressure and manometer pressure
Deflate cuff at rate of 2–3 mm Hg/s
Active or agitated patient Variable Recheck when patient is quiet
Cuff No. (Size) Limb Circumference (cm)
From American Academy of Pediatrics Task Force Pressure Control: Report.
Pediatrics. 1977;59:797, with permission.
58 Section II ■ Physiologic Monitoring
4. Place the stethoscope or Doppler over the brachial
artery for the upper arm or above the popliteal artery for
5. Inflate the cuff rapidly to a pressure 15 mm Hg above
the point at which the brachial pulse disappears.
7. The pressure at which a sound is first heard is the systolic
pressure (Korotkoff I). The pressure at which silence
begins corresponds to the diastolic pressure (Korotkoff V).
The pressure should be measured to the nearest 2 mm Hg.
In patients in whom the sounds do not disappear, the
point at which the sounds change abruptly to a muffled tone
No comments:
Post a Comment
اكتب تعليق حول الموضوع