Chapter 11 ■ End-Tidal Carbon Dioxide Monitoring 77
5. For sidestream capnography, connect the sampling
may introduce an excessive amount of bulk and weight
to the endotracheal tube increasing the risk of tube
2. With sidestream capnography, a low–dead-space
adapter allows for less bulk and weight; however, care
must be taken not to pull excessively on the sample line
that is connected to the measurement instrument (6,7)
Colorimetry provides a quick qualitative measure of CO2 in a
sufficiently fast to detect exhaled CO2 within 1 or 2 breaths.
1. For confirmation of endotracheal tube placement
1. Immediately following endotracheal intubation, attach
T-piece resuscitator or Ambu bag.
2. Within 1 to 2 breaths, the indicator color should
change from purple to yellow with every exhalation if
the tube is within the trachea and not in the esophagus.
Some CO2 detectors have a small plastic strip that
needs to be removed for the gas to flow through.
3. Remove the CO2 detector before attaching the ventilator circuit.
1. This device is not very sensitive when CO2 output is
low, as may be the case in patients with cardiac arrest
and minimal CO2 excretion and in very preterm infants
during initial resuscitation (3,22,23)
1. Walsh BK, Crotwell DN, Restrepo RD. Capnography/
Capnometry during mechanical ventilation: Respir Care. 2011;
2. Galia F, Brimioulle S, Bonnier F, et al. Use of maximum end-tidal
CO(2) values to improve end-tidal CO(2) monitoring accuracy.
4. Rozycki HJ, Sysyn GD, Marshall MK, et al. Mainstream end-tidal
carbon dioxide monitoring in the neonatal intensive care unit.
5. Lightdale JR, Goldmann DA, Feldman HA, et al. Microstream
6. Kugelman A, Zeiger-Aginsky D, Bader D, et al. A novel method of
CO2. Pediatrics. 2008;122(6):e1219.
Fig. 11.3. Infant sidestream low–dead-space adapter with sample
tubing. (Courtesy of Oridion Capnography, Needham, Massachusetts.)
Fig. 11.4. Infant sidestream low–dead-space adapter (arrow) in
No comments:
Post a Comment
اكتب تعليق حول الموضوع