62 Section II ■ Physiologic Monitoring
Ensure that the entire circuit and all the ports are fluid
5. If using disposable transducers, connect the reusable
interface cable to the transducer and to the monitor.
6. Secure the transducer at the patient’s reference level,
defined as the midaxillary line (heart level). If using
transducer holders, level the reference mark on the
holder at the patient’s reference level.
7. Connect the distal end of the circuit to the patient’s
catheter, ensuring that the catheter hub is filled with
8. Start the infusion pump. The pump rate cannot exceed
the flow rate of the flush device.
9. Open the stopcock connected to the transducer to air
(shut off to the patient, open to atmosphere).
10. Zero/calibrate the monitor according to the manufacturer’s instructions.
11. Close the stopcock connected to the transducer (open
12. Set the monitor pressure waveform scale to one that
accommodates the entire pressure wave.
13. Observe the waveform obtained. If the wave appears to
be damped (flattened, poorly defined, with slow rise
time), check the circuit for air bubbles starting at the
Fig. 9.6. Arterial pressure waveforms: normal arterial waveform
(top); dampened arterial waveform (middle); arterial waveform
and not actual pressure values.)
to 4,220 grams. Pediatrics. 1981;67:611, with permission of American Academy of Pediatrics.)
Chapter 9 ■ Blood Pressure Monitoring 63
distal end (Fig. 9.6). If no air bubbles are detected, then
14. Once a stable pressure reading is obtained, set the
alarm limits. Mean arterial pressure value is optimally
used to set alarm limits (Fig. 9.7).
15. Zero the transducer every 8 hours.
16. When blood samples are drawn from the line, flushing
should be done gently with a syringe using a minimal
amount of heparinized saline solution.
2. Cracked Luer-lock connections, causing leaks, low
pressure readings, or blood to back up in the line
4. Malfunctioning infusion pump not providing continuous flush, causing the line to clot off
5. Defective reusable transducer interface cable (disposable transducer system)
6. Erroneous readings caused by the transducer not being
occur when the transducer is lower than the patient’s
7. Problems associated with catheters
8. Tip of the catheter lodging against the wall of the vessel
(will cause the pressure wave to flatten and the pressure
to rise slowly as a result of the continuous infusion)
9. Transducer not zeroed to atmosphere (static pressure
trapped by stopcock valve and a syringe stuck in the
port that should be opened to air). This will cause lower
or negative pressure readings.
10. Loss of blood if stopcock is left open and third port is
11. Fluid overload if a pressurized IV bag is used instead of
an infusion pump and the fast flush mode is used to
Table 9.3 Trouble-shooting for Intravascular Pressure Monitoring
Problem Cause Prevention Treatment
Catheter tip against vessel wall Usually unavoidable Reposition catheter while observing
Partial occlusion of catheter tip by clot Use continuous infusion of normal saline
or ½ normal saline with 0.5–1 U
Remove line if possible. If line
removal is not an option, then
aspirate clot with syringe and
Clotting in stopcock or transducer, or blood in
Flush catheter carefully after blood withdrawal and re-establish continuous
infusion; back-flush stopcocks to
Change in transducer level. A 10 cm change in
height will alter the pressure reading by 7.5
mm Hg. Note: If the cannula is inserted into
the radial artery, raising the hand will not
effect the measurement as long as the transducer is maintained level with the hearta
No comments:
Post a Comment
اكتب تعليق حول الموضوع