3. Cuff inflation will interfere with pulse oximetry measurement in the same limb.
4. Nosocomial infection may result from using the same
cuff for more than one patient.
Oscillometric Measurement of Arterial Blood
Pressure (Automatic Noninvasive)
The oscillometric or noninvasive blood pressure (NIBP)
monitoring technique offers a method for measuring all
arterial blood pressure parameters (systolic, diastolic,
mean, heart rate) (7–15). The underlying principle of this
method is that the arterial wall oscillates when blood flows
in pulsatile fashion through a vessel. These oscillations are
changes (Fig. 9.2). When arterial pressure is just above the
cuff pressure, there is a rapid increase in the amplitude of
the oscillations and this is taken as systolic pressure. The
point at which the amplitude of the oscillations is maximal
technique, the integration of the oscillometric method
within an NIBP algorithm may differ substantially between
1. This technique employs a BP cuff interfaced to a computerized BP monitor.
2. A pneumatic cuff is used in the same fashion as with
3. The monitor employs a miniature computer-controlled
air pump and a bleed valve to control inflation and
4. A pressure transducer interfaced to the cuff tubing
senses the inflation pressure of the cuff and oscillations transmitted to the cuff by the underlying
Fig. 9.1. Cuff of correct size applied to upper arm.
Fig. 9.2. Determination sequence for oscillometric measurement.
Chapter 9 ■ Blood Pressure Monitoring 59
5. The system will inflate the cuff to a level above the
point at which no pulsations are detected.
6. As the cuff is being deflated to the level of the systolic
and pressure oscillations received and transmitted by
7. The systolic pressure is assigned the value of the cuff
pressure at the time oscillations were initially detected.
The diastolic value is determined by the lowest cuff
pressure when there is a sudden decrease in oscillations.
9. Heart rate values are calculated by computing the
mean value of the time interval between pulsations.
10. Higher detection sensitivity allows this technique to be
used on parts of the extremities where auscultatory
methods are not possible (i.e., forearm and lower leg).
1. Measurement of BP in stable infants or when invasive
BP measurement is not required or is unavailable
2. When only intermittent BP measurements are required
1. Severe edema in the limb to be measured; will affect result
2. Decreased perfusion, ischemia, infiltrate or injury in
3. Peripheral venous/arterial catheter in place in limb
1. Provides only intermittent BP measurements
2. Pressure may not be detectable in low-perfusion state
or shock. Do not assume that it is simply an equipment
problem; use clinical correlation.
3. Pressure is not detectable or may be inaccurate in neonates who are restless or having seizures.
4. Inaccurate measurements (Table 9.1)
2. Neonatal cuff (designed for use with the specific
monitor)—cuff may be single-tube or double-tube type,
provided the appropriate adapter is used. Neonatal cuff
sizes range from 1 to 5 (Table 9.2).
1. Incorrect cuff size can significantly alter the BP value
No comments:
Post a Comment
اكتب تعليق حول الموضوع