lead to measurement error, depending on the frequency
response of the capnograph; different capnographs may
have different frequency responses.
3. Contamination of either the monitor or the sampling
system by secretions, blood, or condensation may lead
4. Large dead space affects PetCO2 measurements. The
difference between PetCO2 and PaCO2 increases as
dead-space volume increases and may vary within the
5. The PetCO2 adapter can add to the dead space and
resistance of the respiratory circuit, particularly in small
6. PetCO2 measurements may not provide an accurate
correlation with PaCO2 in small preterm infants with
nonhomogenous lung disease and, therefore, cannot be
substituted for PaCO2 analyses in preterm infants during this critical period (19,20).
7. Acute pulmonary hypoperfusion during cardiac surgery
may be associated with a sudden decrease in PetCO2
and mimic accidental endotracheal extubation (21).
1. Use adaptors specifically designed for neonatal application.
2. For mainstream capnography, an airway adapter is
needed, along with a reusable sensor attachment.
3. For sidestream capnography, an airway adapter with
sampling tube is used (Fig. 11.3).
4. Sidestream technology can be used with nasal prongs
in spontaneously breathing patients.
1. In the mainstream adapter, prevent condensation in
2. In the sidestream adapter, prevent fluid (water) buildup
3. For both mainstream and sidestream, when adding
bulk to the endotracheal tube, extra attention should be
given to properly securing the position of the tube.
4. Tidal volume measurements may be affected if the
PetCO2 adapter is placed between the endotracheal
tube and the ventilator flow sensor.
1. Familiarize yourself with the system before proceeding.
2. Follow manufacturer’s instructions for equipment calibration
3. Attach the adapter in line with the endotracheal tube
and the ventilator T piece (both sidestream and mainstream) (Fig. 11.4).
4. For mainstream capnography, connect the sensor to the
Fig. 11.1. Mainstream capnographic monitor unit and clip-on
sensor assembly. (Courtesy of Respironics, Inc., Murrysville,
Fig. 11.2. Sidestream capnographic monitor unit with sample
line. Note PetCO2 value and wave. (Courtesy of Oridion
Capnography, Needham, Massachusetts.)
No comments:
Post a Comment
اكتب تعليق حول الموضوع