a. An oversized cuff will yield lower BP values; an
undersized cuff will produce higher BP values.
2. Patient must be still during measurements.
3. For optimal infection control, cuffs should be for
4. Oscillometric BP measurement may lose accuracy in
very hypotensive states; this needs to be taken into consideration in such patients (16,17).
1. Become familiar with the monitor and the equipment
to be used. Be aware of the normal BP changes with
gestational and postnatal age (18).
2. Measure the circumference of the extremity where the
cuff is to be applied. Select the appropriately sized cuff
3. Apply the cuff snugly to the limb. The cuff can be
applied over a thin layer of clothing if necessary; however, a bare limb is recommended.
4. Attach the monitor air hoses to the cuff. The limb from
which pressure is to be measured should be level with
5. Turn the monitor on and ensure that it passes the
power-on self-test before proceeding.
6. Press the appropriate button to start a blood pressure
7. If the values obtained from the initial cycle are questionable, repeat the measurement.
Multiple readings with similar values yield the optimal assurance of accuracy.
8. If readings are still questionable after repeating the
cycle, reposition the cuff and repeat the measurement.
9. Periodic inspection of the cuff and extremity is critical
to avoid problems such as cuff detachment or shift in
10. Most NIBP systems can be programmed by the user to
measure BP automatically at user-determined intervals.
The interval between measurements should be long
Fig. 9.3. Oscillometric BP monitor. (Courtesy of GE Healthcare.)
60 Section II ■ Physiologic Monitoring
enough to ensure adequate circulation and minimize
trauma to the limb and skin distal to the cuff.
11. In infants with suspected congenital heart disease,
BP should be measured in all four extremities or at
least the right arm and a leg for pre- and postductal
1. Perfusion in the limb may be compromised if the cuff is
2. Repeated continuous cycling may cause ischemia, purpura, and/or neuropathy in the extremity.
3. Cuff inflation will interfere with pulse oximetry measurement and IV infusion in the same limb.
4. Nosocomial infection may arise from using the same
cuff for more than one patient.
Continuous Blood Pressure Monitoring
Intra-arterial direct continuous BP monitoring is considered
to be the “gold standard” for measuring BP. It has the added
advantage of permitting access for repeated arterial blood
sampling in critically ill neonates.
1. BP measurement is obtained from the vascular system
via a catheter that has been introduced into an artery,
either the umbilical artery in the neonate or a peripheral artery (Chapters 29 and 31).
3. A BP transducer is a device that converts mechanical
forces (pressure) to electrical signals. There are two
a. Strain gauge pressure transducer: Composed of
metal strands or foil that is either stretched or
released by the applied pressure on the diaphragm
(1) Applied pressure causes a proportional and linear change in electrical resistance.
(2) Problems associated with strain gauges include
drift due to temperature changes (departure
from the real signal value), fragility, and cost.
b. Solid-state pressure transducer (semiconductor):
Composed of a silicon chip that undergoes electrical
resistance changes because of the applied pressure
(1) Lower cost, accurate, and disposables
(2) Because of the miniature integration on the
silicone chip, the circuitry necessary to minimize temperature drift is incorporated in the
4. Miniature transducer-tipped catheters are available that
do not depend on fluid-filled lines for the transmission
of pressure. Microtransducer catheters in general have
better fidelity characteristics, but at a much higher cost
than conventional fluid-filled systems and are currently
not generally available for neonatal use
5. The standard medical BP transducer output rating is
5 μV/V/mm Hg. The pressure monitor processes the
kilopascals, including generation of systolic, diastolic,
and mean values. The monitor provides a user-friendly
numerical and graphical display allowing beat-to-beat
measurement of pressure and also allows analysis of the
waveform. Analysis can be clinical (e.g., morphology,
determining the position of the dicrotic notch or
“swing” that can give information regarding filling status and cardiac output) or computerized.
To continuously monitor intravascular pressure
1. In very small or unstable infants, particularly those
with severe hypotension, on inotropic support
2. During major procedures that could cause or exacerbate intravascular instability
3. To monitor infants on ventilator support or extracorporeal membrane oxygenation
4. Allow frequent arterial blood sampling.
None absolute, except for those specific to catheter placement
aorta. Thus, systolic BP in the peripheral arteries can
2. Very small-diameter catheters may result in underreading of systolic BP.
1. Intra-arterial catheter: May be an umbilical arterial
catheter (Chapter 29) or peripheral arterial catheter
2. Pressure monitoring tube: Fluid-filled tubing to couple
arterial cannula to the pressure transducer. This tubing
should be short (not exceed 100 to 120 cm from the
transducer to the patient connection) and stiff (low
No comments:
Post a Comment
اكتب تعليق حول الموضوع