Chapter 9 ■ Blood Pressure Monitoring 61
compliance to reduce damping of pressure wave). A
three-way tap is incorporated in the tubing to allow the
system to be zeroed and blood samples taken.
3. Pressure transducer with cable to signal processor
4. Neonatal physiologic monitor (multiparameter monitoring system)
a. Minimum configuration should have the capability
of displaying systolic, diastolic, and mean pressures
b. It should have provision for high and low alarm settings.
5. Mechanical infusion device (infusion pump) with
syringe and tubing to deliver heparinized saline (0.5 to
1 U heparin/mL of fluid) at 0.5 to 1 mL/h. Pressurized
6. Some disposable pressure-monitoring kits offer closedloop systems for sampling (Fig. 9.5)
a. The system employs a mechanism for aspirating and
holding a fixed amount of blood in the pressure tubing rather than in a syringe.
b. The distal end is equipped with a small chamber
with a rubber septum that allows a self-guiding short
blunt syringe adapter to penetrate and aspirate blood
c. The initial volume pulled back is sufficient to
and is not diluted by the fluid being infused. The
absence of stopcocks at the distal end eliminates a
possible site for contamination. In addition, the
blood pulled back is conserved, and the amount of
fluid used to flush the sample line is reduced.
For catheter placement, see Part 5 of the book, “Vascular
1. Familiarize yourself with the bedside monitor and the
pressure zero/calibration procedure. To maintain
is done in several ways depending on the particular
length of pressure tubing, <12 inches in length,
inserted between the catheter and the pressure
b. Ensure that all the Luer-lock connections are tight
c. If possible, avoid the use of IV tubing components
in the pressure-monitoring circuit.
the infusion pump supplying the flush device should
be set to 0.5 to 3 mL/h and should never exceed the
flow rating of the flush device. When pump flow
1 mL/h is recommended for most arterial lines.
4. For circuit priming, use the solution that will be used
for the continuous infusion. Prime the circuit slowly to
avoid trapping air bubbles in the flush device inlet.
Fig. 9.4. Representative disposable BP transducer setup. (1)
Fig. 9.5. Disposable blood pressure transducer setup with
closed loop system for sampling. A: Umbilical arterial catheter.
B: Stopcock with special valve to draw blood samples. C: High
pressure tubing. D: Stopcock attached to heparinized saline flush
No comments:
Post a Comment
اكتب تعليق حول الموضوع