Chapter 8 ■ Cardiorespiratory Monitoring 51
6. To avoid skin damage, do not use fingernails to remove
7. Secure the patient cable to the patient’s environment to
8. Use only monitors that have been checked for safety
and performance—usually indicated by a dated sticker
9. Do not use monitors with defects such as exposed wires,
broken or dented casing, broken knobs or controls, or
10. Monitor alarms should prompt immediate patient
a. Note alarm indication (i.e., tachycardia or bradycardia).
b. Treat patient condition as necessary or correct the
c. If alarm is silenced or deactivated during the course
of patient evaluation, it should be reactivated prior
to leaving the patient’s bedside.
1. Familiarize yourself with the monitor prior to beginning.
2. Electrode and lead wire placement: Although you
should refer to the monitor manufacturer’s placement
instructions, general electrode placement guidelines
a. Skin preparation: Skin should be clean and dry to
provide the best electrode-to-skin interface.
(1) Wipe skin with an alcohol pad and allow to dry
(2) Avoid the use of tape to secure electrodes—for
optimal performance and proper electrical
interface, electrodes must adhere directly to
(1) Right arm (white): Right lateral chest at level of
(2) Left arm (black): Left lateral chest at level of the
(3) Left leg (red or green): Left lower rib cage
(4) Although this configuration allows the use of the
same electrodes to monitor both ECG and respiration, optimal ECG signal may be obtained
when the right arm lead is at the right midclavicle and the left leg lead is at the xiphoid (4).
c. If not using electrodes with integrated wires, place
electrode pads in basic three-lead configuration as
above, then connect lead wires via electrode clips.
(1) White lead (right arm) to right chest electrode
(2) Black lead (left arm) to left chest electrode
(3) Red or green lead (left leg) to left lower rib cage
3. Turn monitor on—most monitors will conduct an automatic self-test.
4. Connect the patient cable to the monitor.
5. Select the lead that provides the best signal and QRS
size (lead II is usual default) (Fig 8.4).
a. Ensure that heart rate correlates to QRS complexes
seen on display—make sure that the QRS detector is
not counting high or peaked T or P waves.
6. Verify that low and high heart rate alarms are set appropriately.
a. Irritation from alcohol—may occur with even shortterm application to immature skin
b. Trauma caused by rubbing with excessive vigor during skin preparation
c. Irritation from incorrectly formulated electrode gel
d. Secondary effects of skin breakdown
(1) Cellulitis or abscess formation
(2) Increased transepidermal water losses
(3) Hypo- or hyperpigmented marks at sites of prior
irritation or inflammation (Fig. 8.5)
2. Erroneous readings caused by artifacts (5) (Table 8.1)
(1) Sixty-cycle electrical interference (frequency of
52 Section II ■ Physiologic Monitoring
(2) Interference from other equipment used in the
patient’s immediate environment
(3) Electrical spike may be generated when certain
types of polyvinyl chloride tubing are mechanically deformed by infusion pump devices—spikes
appear as ectopic beats on the monitor (rare) (6).
b. Decreased signal amplitude with motion artifact
c. Poor electrode contact or dried electrode gel
d. Incorrect vectors because of inaccurate lead placement (Fig. 8.6)
e. Inappropriate sensitivity settings
a. Hardware or software failure
a. False alarms (either tachycardia or bradycardia)
resulting from inaccurate interpretation of heart rate
b. Inappropriate alarm parameters for patient
1. Reliable and accurate monitoring of neonatal respiratory activity
a. Trending of respiratory activity over time
2. Assessment and surveillance of critically ill neonates
3. To provide early warning of potentially significant
changes in respiratory rate by identifying respiratory
rates above or below preset alarm limits
A low-level, high-frequency signal is passed through
the patient’s chest via surface electrodes.
Fig. 8.4. Typical ECG tracings: Lead I (top), lead II (middle),
Fig. 8.5. Residual hyperpigmented marks on the extremities
present more than 1 year after application of ECG leads for cardiorespiratory monitoring.
1. Gently clean skin with alcohol wipe and
allow to dry prior to electrode reapplication.
2. Check electrode/cable connections.
monitor for improvement in signal quality.
2. After source of interference is identified,
increase distance between that equipment
and patient while rerouting power cords
3. If above maneuver is unsuccessful, replace
60-Hz interference 1. Follow procedure for poor electrode
3. If 1 and 2 are unsuccessful, try an alternate
No comments:
Post a Comment
اكتب تعليق حول الموضوع