6. The RHELAX RF reagent is free from prozone effect
at RF levels between 10 IU/mL to 2300 IU/mL of RF
7. The RHELAX RF reagent is sensitive to the presence of
IgM RF with heterogeneous specificity.
658 Concise Book of Medical Laboratory Technology: Methods and Interpretations Troubleshooting
Problem: False positive results
1. Plasma is used as a test specimen Only serum must be used for testing
be stored for upto a week at 2–8°C
3. Markedly lipemic, hemolyzed and contaminated serum samples
could produre non-specific results
Avoid using lipemic, hemolyzed and contaminated serum samples for
only clean and dry glass slides must be used for testing
6. Latex particles contaminated with positive control/positive
Care must be taken to see that the latex reagent dropper tip does not
touch the sample or control taken on the slide during dispensing of
concentrations. Ensure that exactly one drop of reagent is dispensed
vertically while dispensing the reagent
10.Dried latex particles observed in the latex reagent:
• During slide test with negative control
• In the dropper of the vial (due to freezing of the latex reagent
• Improper dispensing of the entire reagent from dropper
Immediately after performing the test, transfer the contents of the
reagent dropper back into the reagent vial
Ensure that no reagent is left behind in the dropper. Close the cap
of the reagent vial properly and store it back at 2–8°C
Do not freeze the reagent vial
Problem: False negative results
1. The reagent may be damaged due to microbial contamination or
exposure of extreme temperatures
Performance of the reagents can be verified by using positive control/
uniformly and improve test readability
4. Hemolyzed samples may be used for testing Avoid using hemolyzed samples
5. If the conclusion of false negative result has been arrived by
comparison with another kit, this other kit could be giving a false
Run the test with a third kit to validate the results
6. Decrease in drop size, therby leading to less amount of reagent
Less amount of reagent dispensed gives false negative results at
borderline concentrations. Ensure that exactly one drop of reagent is
the RF present in the sample thereby leading to false negative results
Problem: Positive control giving negative reaction
1. The positive control may have deteriorated due to contamination
or exposure to extreme temperatures
Check the performance of the latex reagent using known positive
samples. If the latex reagent is working then the positive control may
Problem: Delayed agglutination
SLIDE TEST FOR ANTI DEOXYRIBONUCLEOPROTEIN (RHELAX SLE)
(Courtesy: Tulip group of companies)
The presence of autoantibodies to nuclear proteins is a
No comments:
Post a Comment
اكتب تعليق حول الموضوع