Initial RF positivity has been a sensitive predictor for later
joint destruction. Quantified measurement of initial RF
level and especially repeated measurements of RF at regular
intervals seems to add significantly to the prognostic value
of RF in distinguishing between progressive and nonprogressive disease in early RA.
QUANTIA-RF is a turbidimetric immunoassay for
quantitative detection of rheumatoid factors of the IgM
1. QUANTIA-RF activation buffer (R1): Ready-to-use
2. QUANTIA-RF latex reagent (R2): Ready-to-use uniform
suspension of polystyrene latex particles coated with
suitably modified Fc fraction of human IgG.
3. QUANTIA-RF calibrator: Lyophilized preparation of RF
positive serum, which is equivalent to stated amount
of RF on IU/mL basis, when hydrated appropriately.
The QUANTIA-RF calibrator is traceable to the WHO,
International Reference Preparation of Rheumatoid
Each batch of reagents undergoes rigorous quality
control at various stages of manufacture for its specificity,
1. Store the reagents at 2 to 8°C. Do not freeze.
2. The shelf-life of the reagent, activation buffer and
the calibrator is as per the expiry date mentioned
on the respective vial labels.
3. The reconstituted QUANTIA-RF calibrator is stable
for 7 days at 2 to 8°C and 48 hours at 25 to 30°C (RT).
4. The working reagent for QUANTIA-RF can be prepared
by mixing R2 and R1 in the ratio 1:5.
5. The mixed stability of the working reagent (R1+ R2) is
7 days when stored at 2 to 8°C.
QUANTIA-RF is a turbidimetric immunoassay for the
determination of rheumatoid factors and is based on the
principle of agglutination reaction. The test specimen
is mixed with QUANTIA RF latex reagent (R2) and
activation buffer (R1) and allowed to react. Presence of RF
in the test specimen results in formation of an insoluble
complex resulting in an increase in turbidity, which is
measured at wavelength 505 to 578 nm. The increase in
turbidity corresponds to the concentration of RF in the test
1. In vitro diagnostic reagent for laboratory and
professional use only. Not for medicinal use.
2. All the reagents derived from human source have
been tested for HBsAg and HIV antibodies and are
found to be non-reactive. However, handle the
3. Reagents contain 0.1% sodium azide as preservative.
Avoid contact with skin and mucosa. On disposal,
flush with large quantities of water.
4. The reagents can be damaged due to microbial
contamination or on exposure to extreme temperatures. It is recommended that the performance
of the reagents be verified using known controls
5. Gently mix the QUANTIA-RF latex reagent well
before use to disperse the latex particles uniformly to
6. The working reagent should be mixed gently.
7. Do not use vortex mixers for mixing. Gently mix the
reagents and samples during test procedures.
8. As the reagents within lots have been matched,
reagents from different lots must not be interchanged.
9. Calibrators of different manufacturers must not be
used with QUANTIA-RF reagents.
10. The calibration curve must be validated periodically
11. The QUANTIA-RF assay is recommended only
for analyzers with cuvette mode. Though any
semiautomated analyzer with appropriate programing
facility can be used, for best results it is recommended
to use; Quantiamate analyzer. Fully automated
analyzers may be used, provided the reagent has been
12. The procedure mentioned here is based on a
minimum reading volume of 500 µL (0.5 mL). In case
of instruments where minimum volume required for
reading absorbance is 1.0 mL, use double the quantity of reagents and samples mentioned in the test
Specimen Collection and Preparation
No special preparation of the patient is required prior to
specimen collection by approved techniques.
Only serum should be used for testing. Should a delay
in testing occur, store the samples at 2 to 8°C. Samples can
be stored for up to 3 days at 2 to 8°C, provided they are not
contaminated. Do not use hemolyzed, icteric or highly
turbid sera. Turbid or particulate serum samples must be
clarified by centrifugation at 2000 rpm for 15 minutes. Use
the clear supernatant for testing.
Spectrophotometer with 505 to 578 nm wavelength
filters and cuvette mode, stopwatch, well-calibrated
micropipettes, disposable tips, isotonic saline, particulate
free distilled water, test tube rack, incubator/waterbath
set at 37°C, optically clean disposable cuvettes such as
Quantiamate semimicrocuvettes/glass cuvettes.
Note: Though any filter between the wavelengths 505 to
578 can be used, optimum results are obtained with a filter
Bring reagents and specimen to room temperature before
Method for Preparation of RF Calibration Curve
The QUANTIA-RF calibrator must be reconstituted
exactly with 1.0 mL of distilled water, wait for 10 minutes,
gently swirl the vial till the solution attains homogeneity.
Once reconstituted, it is ready to use for preparation of
RF calibration curve. The concentration (S) of RF in the
reconstituted calibrator is as mentioned.
Dilute the calibrator serially as mentioned below for
preparation of calibration curve.
Calibrator dilution No. D1 D2 D3 D4 D5
Isotonic saline - 100 µL 100 µL 100 µL 100 µL
100 µL 100 µL 100 µL 100 µL 100 µL
Conc. of IgA in IU/mL 120 60 30 15 7.5
The above five dilutions of the calibrator including
the highest 120 IU/mL (D1) and lowest 7.5 IU/mL (D5)
concentrations of measuring range must be used for the
preparation of the calibration curve.
Test Procedure for Preparation of Calibration Curve
1. Zero the instrument with distilled water.
2. Pipette 400 µL of QUANTIA-RF activation buffer (R1)
and 100 µL of QUANTIA-RF latex reagent (R2) in the
measuring cuvette. Mix well and incubate for five
Pipette 500 µL of QUANTIA-RF working reagent in
the measuring cuvette. Mix well and incubate for five
3. Add 10 µL of calibrator (D1), mix gently and start the
4. Read absorbance (A1), exactly at 10 seconds, and
absorbance (A2) again at the end of exactly 4 minutes.
5. Repeat steps No. 2 to 4 for each diluted calibrator (D2
to D5) for preparing calibration curve.
6. Calculate DA (A2-A1) for each calibrator (D1 to D5).
Plot a graph of DA versus concentration of RF on the
graph paper provided with the kit.
“The calibration curve” so obtained is valid only for the
same lot of QUANTIA-RF reagents.
For determination of RF concentration in the test
1. Follow steps 2 to 4 as mentioned in the above procedure
for calibration curve using the test specimen in place
2. Calculate ∆A (A2-A1) for the test specimen.
If the ∆A of the test specimen is less than ∆A obtained
for the standard of highest concentration (D1) then
the concentration of IgA in the test specimen can be
determined directly by interpolating ∆A of the test
specimen from the calibration curve.
If the ∆A of the diluted test specimen is higher than
∆A of standard with highest concentration (D1), then the
specimen such as 1:10, 1:20, etc. till the DA of the diluted
test specimen is less than ∆A of ∆1. Then proceed for calculations.
No comments:
Post a Comment
اكتب تعليق حول الموضوع