Problem: False positive results
1. Plasma is used as a test specimen Only serum must be used for testing
be stored for upto a week at 2–8°C
clean and dry glass slides must be used for testing
1. The positive control may have deteriorated due to contamination
or exposure to extreme temperatures
Check the performance of the latex reagent; using known positive
samples, if the latex reagent is working then the positive control may
1. The reagent may be damaged due to microbial contamination or
exposure to extreme temperatures.
Performance of the reagents can be verified by using positive control/
uniformly and improve test readability.
4. Hemolyzed samples may be used for testing. Do not use hemolyzed samples for testing.
such as trauma, myocardial infarction and ischemic heart diseases.
Dilute the specimen and rerun the test.
Carry out a semiquantitative assay to determine the CRP level in
6. If the conclusion of false negative results has been arrived at by
comparision with another kit, other kit could be giving a false positive reaction.
Run the test with a third kit to validate results.
7. Decrease in drop size, thereby leading of less amount of reagent
Less amount of reagent dispensed gives false negative results at
borderline concentrations. Ensure that exactly one drop of reagent is
the RF present in the sample thereby leading to false negative results.
Problem: Positive control giving negative reaction
touch the sample or control taken on the slide during dispensing of
7. CRP levels increase in women who are on oral contraceptives Check the history of the patient
concentrations. Ensure that exactly one drop of reagent is dispensed
vertically while dispensing the reagent
11. Dried latex particles observed in the latex reagent
• During slide test with negative control
• In the dropper of the vial (due to freezing of the latex reagent
No comments:
Post a Comment
اكتب تعليق حول الموضوع