Modern blood gas analyzers are electronic marvels
compared to the methods used for this purpose 20 years
ago. On attaching the sample syringe to the cuvette, they
automatically draw the sample into a heated sampling
chamber with miniaturized electrodes that quickly and
accurately (if properly calibrated) measure pH, PCO2 and
PO2 values. Based on these three measured values, these
Blood Gases and Electrolytes 557
units automatically calculate HCO3, total CO2, percent
oxygen saturation and O2 content, which is based on entry
of the patient’s measured hemoglobin values.
A companion to such units, known as a co-oximeter,
directly measures percent oxygen saturation and
hemoglobin, then accurately calculates oxygen content
and carboxyhemoglobin, a value that reflects the degree
of carbon monoxide in the blood in smoke inhalation
In addition to arterial sampling, critical care specialists
often order blood gas panels in blood drawn through a
central venous line since PO2 and O2 content values of this
blood, when compared against arterial PO2 and O2 content,
enable an estimate of cardiac output, another valuable
service performed by blood gas testing. Such samples are
often collected and run from patients undergoing cardiac
catheterization and the results must be returned while the
patient is still on the table.
These are two broad components to the blood gas panel:
respiratory and metabolic. The values reported are as follows:
¾ pH—This is a logarithmic expression of hydrogen ion
concentration—the acidity of alkalinity of the blood.
The normal human arterial pH is 7.4. Any pH below
this is acid, and any pH above it is alkaline. There is a
narrow range of pH values (7.35 to 7.45) that the human
body and its complicated system of enzyme-supported
system operates within. pH values below 7.0 and above
7.6 are incompatible with life.
¾ HCO3—This value is derived through the blood gas
analyzer’s manipulation of the Henderson-Hasselbalch
Equation. An uncompensated decrease in the HCO3
value causes a decline in pH. An increased HCO3 results
in alkalinization of the blood. Either condition can be
life-threatening. Decreased HCO3 is often the result
of kidney or other major organ failure or uncontrolled
diabetes. Increased HCO3 is more rare and is usually
the result of inappropriate administration of certain
drugs such as some kinds of diuretics or an excess of
¾ PCO2—This value is measured directly by the CO2
electrode. An increased PCO2 is often the result of acute,
chronic or impending respiratory failure, whereas
a decreased PCO2 is the result of hyperventilation
stimulated by a metabolic acidosis or hysteria and
severe anxiety reactions. The normal arterial PCO2 is 40
¾ PO2—The partial pressure of oxygen in the blood is
measured directly by a polarographic O2 electrode. The
normal acceptable range is roughly between 85 and
100. An increased PO2 is usually the result of excessive
oxygen administration that needs to be adjusted
downwards on such results. A decreased PO2 is often the
result of any number of respiratory or cardiopulmonary
AVL COMPACT 2 BLOOD GAS ANALYZER (FIG. 21.1)
The AVL compact series is a marvel of design and function.
As with all AVL blood gas analyzers, this little workhorse is
designed to make operation as simple as possible allowing for
requirements of critical care analysis and is suitable for
intensive care situations, neonatology and lung function
testing. Patient data entry is a breeze via the built-in keyboard
and LCD display. Furthermore, a logical menu guides the user
through all functions making operation simple, reliable and
safe. The simplicity and small footprint of the AVL compact
series mask its advanced analytical performance and data
processing capabilities. Measured parameters in the AVL
compact series include PH, PCO2, PO2, and PBaro. Features
of the AVL compact series include automatic calibration and
cleaning cycles and 20-second analysis time, 32 samples/h.
Other features of the AVL compact series are LCD display
with integrated numeric keypad, automatic sample handling
system, and on-board quality control features able to store up
to 34 QC results for three levels.
FIG. 21.1: AVL compact 2 blood gas analyzer
Dimensions: 14 × 13 × 14 (36 × 33 × 36 cm)
ELECTROLYTE ANALYSIS BY FLAMEPHOTOMETER
In this instrument, the solution under test is passed under
carefully controlled conditions as a very fine spray in the
air supply to a burner. In the flame, the solution evaporates
and the salt dissociates to give neutral atoms. Some of these
though only a very small proportion, move into a higher
energy state (their electrons move to the outer orbits). When
the electrons of these atoms fall back to their original orbits,
they release energy in the form of light. Which is used in
flame photometry of this type and is called emission flame
photometry. Light of characteristic wavelengths is emitted
and passes through a suitable filter, grating or prism on to
a photocell, and the amount of current thus produced is
measured. This varies with the concentration of sodium;
for example, in the solution being tested. Using solutions
of known sodium concentrations, a calibration curve can
be made and this is used for reading the sodium content of
Many gases have been used for the flame. These include
acetylene, propane, butane and coal gas. Both the gas
pressure and air pressure have to be carefully regulated so
as to maintain a constant steady flame, which should be
blue in color and have no yellow streaks. The spray is formed
by passing compressed air through an atomizer, into which
the liquid which is being tested is drawn either by suction
or gravity, and then enters the burner in its air supply. The
pressure used is usually about 10 to 15 lb per square inch.
The atomizer is a very important part of the apparatus.
A steady fine spray of droplets of uniform size should be
produced if there is to be a constant emission of light.
The light produced is first passed through a lens to focus
it and then through suitable filters, a diffraction grating or
prism before it falls on to the photocell. For determining
sodium, an orange or yellow (589 nm) filter can be used.
Potassium emits light at 404.4 and 766.5 nm. For the
former, a violet; and for the latter a deep red filter should
be used. Lithium (671 nm) needs a near red filter for its
The most satisfactory dilution to use should be
established experimentally. For sodium, this may be 1
in 100, 1 in 200, even 1 in 500; for potassium, it is usually
lower, often 1 in 50, but it is possible by varying the
sensitivity to use the same dilution for both. Both sodium
and potassium can interfere with each other, therefore, in
standard solution both the elements are added.
Equipment: Systronics provides digital read-out flame
photometer (Mediflame 127/128/129) that provides two
Microprocessor-based Automation
The microprocessor provides automation in operation,
measurements, and end-result presentation. The unit can
do the estimation of sodium (Na), potassium (K), calcium
(Ca) and lithium (Li) in single aspiration of a sample.
FIG. 21.3: Measurement results can be printed
Blood Gases and Electrolytes 559
For user’s convenience, the unit offers three measuring
modes: (i) serum, (ii) urine, and (iii) bio-fluid, general. The
last mode is helpful for analyzing biological samples other
Frequently used measurement set-ups can be stored
once (in a battery-back-up memory) and recalled
whenever required with a stroke of a button. This
eliminates the typical chores of instructions required to
be given to a microprocessor-based instrument before it
starts the operation. Facility for restandardization with a
single standard is available to minimize the effect of any
unforeseen drift without going for the full recalibration.
A 4-line 20-character LCD readout provides easy user
interface and presentation of results. A Centronics printer
port is provided for Epson compatible dot matrix/inkjet and
HP compatible laser printers. Printer is optional (Fig. 21.3).
¾ Specifically designed for medical applications
¾ Up to four elements measured with single aspiration
¾ Automatic filter selection (narrow band interference)
¾ Curve fittings for nonlinear range (up to 5 standards)
¾ Calibration stored in memory
¾ Restandardization cuts on full recalibration
¾ Record kept of date and time of analysis
¾ Saved set-ups cut operation steps
¾ Measurement results can be recalled later on for display
¾ Measurement results can be printed (individual, full
Batch, of the day, all in memory)
¾ 4-line, 20-character alphanumeric LCD readout
¾ Centronics printer port for epson compatible dot
matrix/inkjet and HP laser printers
¾ Compressor with built-in air filter and air regulator.
Specifically designed for medical use.
Step-by-step guidance for operation.
Digital monitoring of flame stability.
A restandardization facility cuts on full recalibration
Measurement results can be recalled later on for display.
Saved set-ups drastically cut operation steps.
Element Serum Urine BIP-Fluids
Na 100–200 mEq/L 0–250 mEq/L All the four
1:100 dil 1:100 dil elements up
K 0–10 mEq/L 0–100 mEq/L 250 mEq/L
1:100 dil 1:100 dil with 1:100 dil
i. Curve fit software is provided for urine and biofluids.
ii. Curve fit accuracy: ± 2% f.s.
iii. Suitable dilution for concentrations higher than given in the
¾ Filters (10 nm typical): Na and K supplied; Li and Ca
¾ Minimum sample: Approximately 3 mL per element (at
¾ Averaging: 2 to 15 seconds, selectable
¾ Operating air pressure: 0.45 kg/cm2
¾ Air compressor: With built-in air regulator and air filter
to deliver stable and moisture/oil free air supply
¾ Power Supply: 230 Vac +/– 10%, 50 Hz.
System: The system consists of main unit with one each of
Na and K narrow band filters, a compressor with built-in
air regulator and air filter. Li and Ca filters and printer are
No comments:
Post a Comment
اكتب تعليق حول الموضوع