The “Hook Effect” is characterized by the production
of artefactually low results from samples that have
extraordinarily high concentrations of antigen (analyte), far
exceeding the concentration of the upper standard in the
The hook effect is most commonly found in single-step
immunometric assays, a popular format, chosen for its
specificity and speed, particularly with high-throughput
immunoassay analyzers. The assays most affected are
those that have analyte concentration that may range
over several orders of magnitude. For example, alpha
fetoprotein (AFP), CA-125, hCG, PSA, TSH, prolactin and
ferritin are most affected by Hook effect.
The incidence of Hook effect can be reduced (but not
eliminated) by careful assay design—incorporating a wash
step prior to addition of the second antibody, thereby
avoiding simultaneous saturation of both antibodies.
effect by careful assay design, the only reliable method of
routinely eliminating the effect is to test the samples that
are likely to be affected by Hook effect in undiluted and
also at a suitable dilution. Such samples should be diluted
using either the assay diluent or serum from a normal
subject until a stable quantitative response is achieved.
Sometimes with ELISA performed in a microwell plate
unexpectedly higher (or lower) optical densities (OD)
are measured in the peripheral wells than in the central
wells. This phenomenon is called “edge effect”. The
most probable causes of this effect are illumination or
temperature differences between the peripheral and the
Light may cause edge effect if the substrate is
photosensitive (i.e. converted by light exposure) like the
H2O2/OPD substrate in the peroxidase system. Thus, if
strong light is coming from one side (e.g. sunlight from
a window) during the substrate reaction, the peripheral
wells closest to the light source may give elevated OD
values. Temperature difference, however, is the most
Incubation at 37°C instead of room temperature is often
used for shortening incubation time, which is not correct.
Also, a common mistake is to use reactant liquids straight
from a refrigerator and then incubate in a 37°C incubator
(or at room temperature). Temperature changes of these
magnitudes may, especially with short incubation times,
destroy the assay homogeneity in microwell plates. The
peripheral wells will normally be heated up first because of
their position closest to the lower edge of the plate, which
is in direct contact with the warm incubator shelf, which
may result in higher OD values in these wells, other things
being equal. The edge effect may be more pronounced if
plates are stacked during incubation, especially in plates
in the middle of the stack because their central wells are
shielded from the warmer surroundings by the plates
To avoid the above-mentioned problems, the following
¾ Incubations should take place in subdued light or in
the dark (if protocol requires)
¾ Reactant liquids (and plates) should be adjusted to the
temperature intended for incubation
¾ Plates should be sealed with adhesive tape or placed
in a 100% relative humidity environment during
It is one of the most important requirements of
immunoassays. Interference occurs in all situations in
which the antibody is not absolutely specific for the analyte.
Consequently, assessment of specificity is a vital step in the
optimization of every new immunoassay. Poor specificity
results in interference from compounds of similar molecular
structure or which carry similar immunoreactive epitopes.
In determining the overall specificity of an assay, a major
factor is the crossreactivity of the antibody.
Some of the major specificity problem areas are related
to measurement of steroids and structurally related
compounds. All commonly used testosterone assays, cross
react in varying degrees with 5 α-dihydrotestosterone, and
all cortisol assays cross react with prednisolone.
Assessment of the specificity of immunometric assays
employed, each having unique specificity for a different
epitope on the antigen. It is usual practice to employ at
least one monoclonal antibody, which can be selected by
epitope mapping to react only with predetermined sites on
the antigen molecule. Use of two monoclonal antibodies
can introduce extreme specificity.
The ability of a kit to detect very low concentrations of an
analyte (in quantitative ELISA) is mainly understood by
the sensitivity of the kit. Many manufacturers mention the
sensitivity and specificity after the result interpretation.
This is overlooked commonly. One should observe this
carefully. Higher sensitivity is a desirable property in
any kit. Some doubts have been expressed regarding the
value of ultrasensitive assays, which detect very minute
amounts of analyte, which may be below the clinically or
diagnostically significant values.
Most diagnostic kits are not exhausted overnight.
Repeated usage and storage exposes the kit to multiple
thermal shocks. This affects the performance of the kit
over a period of time due to lowering of sensitivity. This
shift in sensitivity affects the ultrasensitive kits lesser than
A good example of ultrasensitive kit is “Third Generation
TSH kits” which are very useful in the diagnosis of
As compared to low sensitive kits, ultrasensitive kits are
more robust, more accurate that improve the reliability
of results and provide confidence to the clinicians on the
CHEMILUMINESCENCE: THE TECHNOLOGY
“Chemiluminescence” is defined as the production of
electromagnetic (ultraviolet, visible or near-infrared)
radiation as a result of a chemical reaction. One of the
reaction products is in an excited state and emits light on
returning to its ground state.
The generation of signal and its estimation varies from
technology to technology. In RIA (radioimmunoassay)
the radioactive signal is measured in gamma counter. In
ELISA, the enzyme and substrate react to produce color,
which is measured using an ELISA reader. Fluorescence
immunoassays involve a similar principle where enzyme
and substrate react to produce a fluorophor, which is
measured fluorometrically. In case of chemiluminescence
immunoassays, the light is produced which is measured.
Measurement of light from a chemical reaction is
highly useful because the concentration of unknown
can be inferred from the rate at which light is emitted.
The rate of light output is directly related to the amount
of light emitted. This type of luminescence is frequently
compared with fluorescence, which also involves emission
of light as a result of relaxation of excited states. Since,
chemiluminescence does not involve initial absorption
of light, measurement of chemiluminescence emission
are made against a lower background noise that is not
possible with conventional fluorescence, thus potentially
and the ability to easily measure very low and very high
light intensities with simple instrumentation provide a
large potential dynamic range of measurement. Linear
measurement over a dynamic range of 106
purified compounds and standards has become possible
with developments in the technology.
Light, as we see it, consists of billions of tiny packets
of energy called photons, which are measured in the
detection process. There are different factors that affect
the emission and measurement of light.
quantum yield, ÖCL, which describes the number of
moles of photons emitted per mole of reactant
¾ The quantity of signal required to produce the emission
¾ Instrumentation employed for the quantification of
Components of Chemiluminescent System
The signal (or substrate) used for generation of light
should have optimum stability. There are many signal
reagent available—luminol, 1,2 Dioxetanes, Acridinium
ester, ruthenium salts, etc. Luminol is preferred of all these
because of its stability and its advantage of being enhanced
No comments:
Post a Comment
اكتب تعليق حول الموضوع