4. A decrease in thyrotropin values has been reported
with the administration of propranolol, methimazol,
5. Genetic variations or degradation of intact TSH into
subunits may affect the biding characteristics of the
antibodies and influence the final result. Such samples
* The data presented above is for illustration only and should not
be used in lieu of a dose response curve prepared with each assay.
In addition, the RLU’s of the calibrators have been normalized to
approximately 100,000 RLU’s for the A calibrator (greatest light output).
This conversion eliminates differences caused by efficiency of the
various instruments that can be used to measure light output.
FIG. 22.24: Example showing average RLU intersects calibration curve
Note 1: Computer data reduction software designed for
chemiluminescence assays may also be used for the data
reduction. Duplicates of the unknown may be averaged as
Note 2: Monobind can assist the laboratory in the purchase
and implementation of equipment/software to measure
and interpret chemiluminescence data.
1. The Dose Response Curve should be within established parameters.
2. Four out of six quality control pools should be within
1. It is important that the time of reaction in each well
is held constant for reproducible results. Pipetting of
samples should not extend beyond ten (10) minutes
to avoid assay drift. If more than one (1) plate is used,
it is recommended to repeat the dose response curve.
Failure to remove adhering solution adequately in the
aspiration or decantation wash step(s) may result in
poor replication and spurious results.
2. Sample(s), which are contaminated microbiologically,
should not be used in the assay. Highly lipemeic or
hemolyzed specimen(s) should similarly not be used.
3. Patient specimens with TSH concentrations above 40
µIU/mL may be diluted with the zero calibrator and
reassayed. The sample’s concentration is obtained by
multiplying the result by the dilution factor.
4. Each component in one assay should be of the same
lot number and stored under identical conditions.
1. If computer controlled data reduction is used to
interpret the results of the test, it is imperative that the
predicted values for the calibrators fall within 10% of
2. Serum TSH concentration is dependent upon a
multiplicity of factors: Hypothalamus gland function,
thyroid gland function, and the responsiveness of
pituitary to TRH. Thus, thyrotropin concentration alone
is not sufficient to assess clinical status.
3. Serum TSH values may be elevated by pharmacological intervention. Domperiodone, amiodazon,
iodide, phenobarbital, and phenytoin have been
reported to increase TSH levels.
assay systems due to the reactivity of the antibodies
“Not intended for newborn screening”
A study of euthyroid adult population was undertaken to
determine expected values for the TSH CIA Microplate
Test System. The number and determined range are given
in Table 22.2. A nonparametric method (95% Percentile
It is important to keep in mind that expected values
for normal population is dependent upon a multiplicity
of factors: The specificity of the method, the population
tested and the precision of the method in the hands of
the analyst. For these reasons each laboratory should
depend upon the range of expected values established
by the Manufacturer only until an in-house range can
be determined by the analysts using the method with a
population indigenous to the area in which the laboratory
TABLE 22.2: Expected values for the TSH CIA test system (in µIU/mL)
70% confidence intervals for 2.5 percentile
of antigen, prepared from normal tissues, most commonly
beef heart. VDRL slide test can be used both qualitatively
and quantitatively for the detection of “Reagin” in serum.
As other better methods are available, this method is hardly even used.
A phospholipid viz. cardiolipin, derived from beef heart
muscle together with cholesterol and lecithin, is used as
an antigen. After mixing the antigen with patient’s serum,
the reaction is accelerated by rotatory agitation either
on a mechanical shaker or by hand. The antigen reacts
with reagin and forms floccules. These floccules can be
observed with naked eye, hand lens or under a low power
objective of a microscope when the reaction is weak.
1. VDRL Test Slide: A 2” × 3” glass slide with 12 paraffin
or ceramic rings of approximately 14 mm inside
2. Hypodermic needles without bevels (18, 19 gauge).
4. Thirty mL flat or concave inner bottomed glass
stoppered, narrow mouth bottle, approximately 35
mm in diameter (bottles with convex inner bottom
1. Use clean and dry glassware.
2. Allow all reagents and samples to reach room
temperature before starting the test.
3. Carry out the test at room temperature (preferably
No comments:
Post a Comment
اكتب تعليق حول الموضوع