(Courtesy: Tulip Group of Companies)
Syphfinal reagent is a ready-to-use, uniform suspension
of polystyrene latex particles coated with cardiolipin,
suspended in a suitable buffer of proprietary composition.
Though Syphfinal reagent, in performance, corresponds
to the other USRs, it accords better readability to the test
results, thereby giving better confidence in reporting
results. As against the conventional VDRL tests, the
samples do not require heat inactivation.
Each batch of reagent undergoes rigorous quality
control at various stages of its manufacture for its
sensitivity, specificity and performance.
Store the reagent at 2–8°C. Do not freeze. The shelf life of the
reagent is as per the expiry date mentioned on the reagent
vial labels. Avoid exposure to elevated temperatures, air
and direct sunlight as the reagents are highly sensitive to
denaturation, drying and microbial contamination.
Syphfinal, latex VDRL reagent, is based on the principle
of agglutination. The test specimen, serum or plasma is
mixed with Syphfinal latex VDRL reagent and allowed to
react for 6 minutes. If antilipoidal antibodies are present in
the specimen, they will react with the latex reagent forming
a visible agglutination. If antilipoidal antibodies are not
present in the specimen, then no agglutination is observed.
616 Concise Book of Medical Laboratory Technology: Methods and Interpretations Note
1. In vitro diagnostic reagent for laboratory and professional
use only. Not for medicinal use.
2. The reagent contains sodium azide 0.1% as preservative.
Avoid contact with skin and mucosa. On disposal,
flush with large quantities of water.
3. All reagents derived from human source are tested
for HBsAg and anti-HIV antibodies and found to
be nonreactive. However, handle the material as if
4. Syphfinal latex VDRL reagent should be gently but
thoroughly mixed before testing to disperse the latex
particles uniformly and improve test readability.
5. Performance of the reagents must be verified with
positive and negative controls provided with the kit
and it is recommended that controls be run with each
No special preparation of the patient is required prior to
sample collection by approved techniques. Do not use
hemolyzed samples. Fresh serum or plasma should be
used for testing. Samples not tested immediately may be
stored at 2 to 8°C for up to 48 hours. Hazy samples should
be centrifuged. Use clear supernatant for testing.
Material Provided with the Kit
Latex VDRL reagent, positive control reactive with the reagent,
negative control nonreactive with the reagent, disposable
slides with eight reaction circles, disposable sample/control
dispensing pipettes, mixing sticks, rubber teats.
Stopwatch, high intensity direct light source, isotonic
Bring all reagents and samples to room temperature
1. Pipette one drop of test sample onto one of the
reaction circles of the disposable slide using a sample
dispensing pipette. The disposable slides and the
sample dispensing pipettes are provided with the kit.
2. Repeat the procedure with positive and negative
3. Add one drop of well-mixed Syphfinal latex reagent to
the test sample, positive control and negative control
respectively. Do not let the dropper tip touch the liquid
4. Using a mixing stick, mix the test sample and Syphfinal
reagent thoroughly spreading uniformly over the
5. Immediately start a stopwatch. Rotate the slide
gently and continuously, observing for agglutination
1. Using isotonic saline prepare serial dilutions of the test
sample, e.g. 1:2, 1:4, 1:8, 1:16, 1:32, 1:64, 1:128, etc.
2. Perform qualitative test procedure using each dilution
3. The titre is reported as the reciprocal of the highest
dilution which shows a positive test result.
Agglutination is a positive test result and indicates the
presence of antilipoidal antibodies in the test sample.
No agglutination is a negative test result indicating the
absence of detectable levels of antilipoidal antibodies in
The titre of antilipoidal antibodies is the highest dilution of
the test sample giving a positive test result (i.e. agglutination).
1. Quantitative procedure must be performed to
determine the response to treatment and detect
2. False positive reactions occur not infrequently and
have been attributed to a variety of acute and chronic
3. In the absence of supporting clinical, historical or
epidemiological evidence, reactive results must be
confirmed with more specific Treponema tests.
4. It is recommended that the results of the test should
be correlated with the clinical findings to arrive at the
5. Dispose all used and contaminated materials as per
Standard Biohazard Safety Guidelines.
Problem: False positive results
1. Reagent is contaminated because of unclean dropper and hence
not suitable for subsequent use
The reagent dropper provided for dispensing the reagent should be
Care must be taken to ensure that the reagent dropper tip does not
touch the liquid on the slide while dispensing
2. Drying of the reagent on slide Do not interpret test results beyond 6 minutes. Testing should not
be carried out under the fan or under conditions where drying is enhanced
3. False positive reactions can also be attributed to a variety of acute
Check the history of the patient. The test result must be correlated
4. False positives may occur due to overspill from one circle to
Care must be taken to see that there is no overspill during rotation
1. Negative control contaminated with positive control/positive
and should not be used for further testing
Problem: Negative control giving positive reaction
Problem: False negative results
1. The reagent may be damaged by exposure to elevated
temperatures, air and direct sunlight, as it is highly
sensitive to denaturation, drying and microbial contamination
Check the performance of the reagent with positive and negative controls provided
with the kit. Avoid exposure of the reagent to light. The vial must be closed properly after use
4. Excess sample dispensed/reagent dispensed leading
5. Hemolyzed or lipemic samples may have been used
Avoid using hemolyzed or lipemic samples for testing
6. Reagent is contaminated because of unclean dropper
and hence not suitable for subsequent use
The reagent dropper provided for dispensing the reagent should be thoroughly
CARBON ANTIGEN FOR SYPHILIS TESTING
(Courtesy: Tulip Group of Companies)
The Rapid Plasma Reagin (RPR)/Carbon Antigen test is
a macroscopic non-Treponema flocculation test for the
detection and quantitation of antilipoidal antibodies.
Non-Treponema tests like CARBOGEN are of great value
when used for screening and follow-up of therapy.
1. CARBOGEN reagent:Aparticulate carbon suspension
2. Positive control,reactivewiththeCARBOGENreagent.
No comments:
Post a Comment
اكتب تعليق حول الموضوع