Note: Do not use the working substrate if it looks blue.
Before proceeding with the assay, bring all reagents,
serum references and controls to room temperature
1. Format the microplates’ wells for each serum
reference, control and patient specimen to be assayed
in duplicate. Replace any unused microwell strips
back into the aluminum bag, seal and store at 2–8°C.
2. Pipette 0.050 mL (50 µL) of the appropriate serum
reference, control or specimen into the assigned well.
3. Add 0.100 mL (100 µL) of the TSH Enzyme Reagent
to each well. It is very important to dispense all
reagents close to the bottom of the coated well.
4. Swirl the microplate gently for 20–30 seconds to mix
5. Incubate 60 minutes at room temperature.**
6. Discard the contents of the microplate by decantation
or aspiration. If decanting, tap and blot the plate dry
7. Add 300 µL of wash buffer (see Reagent Preparation
Section) decant (tap and blot) or aspirate. Repeat two
(2) additional times for a total of three (3) washes.
An automatic or manual plate washer can be used.
Follow the manufacturer’s instruction for proper
usage. If a squeeze bottle is employed, fill each well
by depressing the container (avoiding air bubbles) to
dispense the wash. Decant the wash and repeat two
8. Add 0.100 mL (100 µL) of working substrate solution
to all wells (see Reagent Preparation Section). Always
add reagents in the same order to minimize reaction
time differences between wells.
Do not shake the plate after substrate addition.
9. Incubate at room temperature for fifteen (15)
10. Add 0.050 mL (50 µL) of stop solution to each well
and mix gently for 15–20 seconds. Always add
reagents in the same order to minimize reaction
time differences between wells.
11. Read the absorbance in each well at 450 nm (using
a reference wavelength of 620–630 nm to minimize
well imperfections) in a microplate reader. The
results should be read within thirty (30) minutes of
** For better low-end sensitivity (< 0.5 µIU/mL). Incubate
120 minutes at room temperature. The 40 µIU/mL
calibrator should be excluded since absorbance over 3.0
units will be experienced. Follow the remaining steps.
Each laboratory should assay controls at levels in the low,
normal, and high range for monitoring assay performance.
These controls should be treated as unknowns and values
determined in every test procedure performed. Quality
control charts should be maintained to follow the
performance of the supplied reagents. Pertinent statistical
methods should be employed to ascertain trends. The
individual laboratory should set acceptable asssay
performance limits. Other parameters that should be
monitored include the 80, 50 and 20% intercepts of the
dose response curve for run-to-run reproducibility. In
addition, maximum absorbance should be consistent with
past experience. Significant deviation from established
performance can indicate unnoticed change in experimental
conditions or degradation of kit reagents. Fresh reagents
should be used to determine the reason for the variations.
A dose response curve is used to ascertain the concentration
of thyrotropin in unknown specimens.
1. Record the absorbance obtained from the printout
of the microplate reader as outlined in following
example (An example of the 120-minute incubation
2. Plot the absorbance for each duplicate serum reference
versus the corresponding TSH concentration in
µIU/mL on linear graph paper (do not average the
duplicates of the serum references before plotting).
3. Draw the best-fit curve through the plotted points.
4. To determine the concentration of TSH for an
unknown, locate the average absorbance of the
duplicates for each unknown on the vertical axis of
the graph, find the intersecting point on the curve,
and read the concentration (in µIU/mL) from the
horizontal axis of the graph (the duplicates of the
unknown may be averaged as indicated). In the
following example, the average absorbance (1.019)
intersects the dose response curve at (15.3 µIU/mL)
TSH concentration (Fig. 22.23).
In order for the assay results to be considered valid the
following criteria should be met:
1. The absorbance (OD) of calibrator 0 ng/dL should be
2. Four out of 6 quality control pools should be within
*The data presented above are for illustration only and should not
be used in lieu of a dose response curve prepared with each assay.
FIG. 22.23: Example showing average absorbance intersects dose
response curve at TSH concentration
1. It is important that the time of reaction in each well is
held constant for reproducible results.
2. Pipetting of samples should not extend beyond ten
(10) minutes to avoid assay drift.
3. If more than one (1) plate is used, it is recommended
to repeat the dose response curve.
4. Addition of the substrate solution initiates a kinetic
reaction, which is terminated by the addition of the stop
solution. Therefore, the addition of the substrate and the
stopping solution should be added in the same sequence
to eliminate any time-deviation during reaction.
5. Plate readers measure vertically. Do not touch the
6. Failure to remove adhering solution adequately in the
aspiration or decantation wash step(s) may result in
poor replication and spurious results.
7. Use components from the same lot. No intermixing of
reagents from different batches.
8. Highly lipemic, hemolyzed or grossly contaminated
specimen(s) should not be used.
1. If computer controlled data reduction is used to
interpret the results of the test, it is imperative that the
predicted values for the calibrators fall within 10% of
2. Serum TSH concentration is dependent upon a
multiplicity of factors: Hypothalamus gland function,
thyroid gland function, and the responsiveness of
pituitary to TRH. Thus, thyrotropin concentration
alone is not sufficient to assess clinical status.
3. Serum TSH values may be elevated by pharmacological
intervention. Domperiodone, amiodazon, iodide,
No comments:
Post a Comment
اكتب تعليق حول الموضوع