produce a signal in the measuring system.”

Assay interference can be “analyte dependent or analyte

independent” and can increase or decrease the measured

result.

Increase (positive interference) is due to lack of specificity.

Decrease (negative interference) is due to lack of

sensitivity.

Assay interference can be of different types:

¾ Preanalytical errors

¾ Analytical errors

¾ Postanalytical errors (Fig. 22.18).

Preanalytical Variables

All factors associated with the procedures before the actual

performance of the test are known as preanalytical errors.

They can be as follows:

Patient Based

Such as incorrect sampling times and environmental

factors such as smoking, etc. may change analyte

concentration and consequently interpretation.

Specimen Based

There are many factors that constitute this.

Blood collection procedure and time of collection.

Certain hormones are affected by the time of collection.

Nature of the Sample

For all immunoassays, serum is the matrix of choice.

Samples collected in to tubes containing sodium fluoride

may be unsuitable for some enzymatic immunoassay

methods; preservation with sodium fluoride may affect

results. Impurities in tracers interfere with direct dialysis

methods for free hormones.

FIG. 22.17: Analyzers (Courtesy: Lilac Medicare)

AE 600 Semi-auto analyzer

ELDEX 3.8 Strip reader

580 Concise Book of Medical Laboratory Technology: Methods and Interpretations Hemolysis and Hyperbilirubinemia

Lipemia may cause interference with assays for fat-soluble

compounds such as steroids.

Stability and storage of reagents and samples are as

follows:

Assay Based

Certain procedures before the test is performed like:

Bringing the kit to the room temperature

Checking the incubator temperature

Checking the room temperature

Formatting and arranging the workbench

Planning for assays to be performed

Proper dilution calculation

Maintaining the fridge temperature

Dilution of samples with distilled waterinstead of zero

calibrator.

Manufacturing error:

Batch problem

Packing error

Assay protocol error

Labeling error

Wrong pack insert.

Analytical Variables

These form the major part of all errors that affects the

results in immunoassays. Mostly, these are overlooked and

so attention is not paid to rectify them. These will lead to

erroneous results. They are mainly the procedural errors.

They are mostly confined to the different steps—addition,

washing, incubation, dilution, pipetting, reading the

protocol suggested by the manufacturer is not followed.

Some of the errors are as follows:

Washing Errors

¾ High pressure washing

¾ Carry over during washing

¾ Drying of wells

¾ Splashing during washing

¾ Non-removal of all wash solution from wells or tubes

¾ Not following soak time (if present) during washing

¾ Using a syringe to wash

¾ Leaving bubbles in the well after washing

¾ Not tapping the well after washing

¾ Use of contaminated wash buffer

¾ Wells/tubes falling off while washing.

FIG. 22.18: Factors affecting EIA

Serology/Immunology 581

Pipetting Error

¾ Reuse of pipette tips

¾ Pipette tip blocked

¾ Pipette/dispenser not primed

¾ Pipette barrel contaminated

¾ Using tips to break bubbles

¾ Using same tips to break bubbles.

Equipment Error

¾ Incubators not maintained at right temperature

¾ Heating not uniform throughout

¾ Washer probes blocked, contaminated

¾ Refrigerator not maintaining right temperature

¾ Defrost water falling on kits

¾ Use of dry incubators instead of water bath

¾ Instrument filters not checked periodically.

Procedural Errors

¾ Interchange of reagent lots

¾ Wells not covered during incubation

¾ Not blanking when required

¾ Not running all calibrators to plot a graph

¾ Reuse of pipette tips

¾ Bubbles in wells

¾ Use of contaminated or uncleaned tubes to prepare

reagents

¾ Using kits/reagents beyond expiry

¾ Using negative wells again

¾ Not mixing after adding stop solution.

Postanalytical Variables

The errors that occur after the performance of the test are

called as postanalytical errors. These are like:

¾ Calculation mistakes

¾ Choosing a wrong graph scale

¾ Comparison of result with inappropriate reference

interval

¾ Not correlating the results with clinical history

¾ Transcription error when report is prepared.

Use of Wrong Reference Values

The reference ranges (normal ranges) of various parameters

are different. Manufacturers specify the reference ranges.

The units for reference ranges may differ. One should only

compare with reference intervals given in the pack insert

or should establish their own reference interval.

Comparison with results of different reference intervals

will create confusion. Many laboratories make the mistake

of comparing results with other labs without knowing the

assay conditions and other factors that affect ELISA.

Use of Wrong Units

The units given by manufacturers may be different. For

example, T3 ng/dL is different from ng/mL. A kit having

a sensitivity of 0.4 ng/dL is more sensitive than 0.4 ng/mL

(it is 40 ng/dL).

One should always make note of the units. Reports from

different laboratories may differ in this aspect and create

confusion.

The errors may be of any kind, but the outcome is that

the result is incorrect and hence, a wrong report is given.

To overcome this, one should follow all the steps and

adhere to the protocol strictly.

ELISA Troubleshooting

ELISA is a technique of multiple steps. The steps must

be followed strictly to achieve good results. Errors at any

levels will affect the final result. Complete understanding

of the process is necessary for troubleshooting.

Practical Tips on ELISA

Factors affecting EIA are shown in Figure 22.18.

Normal Washing

In washing plate manually, the most important factor is

that each well receives the washing solution so that, no air

bubbles are trapped in the well or a thumb is not placed

over corner wells.

Comments

Search This Blog

Archive

Show more

Popular posts from this blog

TRIPASS XR تري باس

CELEPHI 200 MG, Gélule

ZENOXIA 15 MG, Comprimé

VOXCIB 200 MG, Gélule

Kana Brax Laberax

فومي كايند

بعض الادويه نجد رموز عليها مثل IR ، MR, XR, CR, SR , DS ماذا تعني هذه الرموز

NIFLURIL 700 MG, Suppositoire adulte

Antifongiques مضادات الفطريات

Popular posts from this blog

علاقة البيبي بالفراولة بالالفا فيتو بروتين

التغيرات الخمس التي تحدث للجسم عند المشي

إحصائيات سنة 2020 | تعداد سكَان دول إفريقيا تنازليا :

ما هو الليمونير للأسنان ؟

ACUPAN 20 MG, Solution injectable

CELEPHI 200 MG, Gélule

الام الظهر

VOXCIB 200 MG, Gélule

ميبستان

Popular posts from this blog

TRIPASS XR تري باس

CELEPHI 200 MG, Gélule

Popular posts from this blog

TRIPASS XR تري باس

CELEPHI 200 MG, Gélule

ZENOXIA 15 MG, Comprimé

VOXCIB 200 MG, Gélule

Kana Brax Laberax

فومي كايند

بعض الادويه نجد رموز عليها مثل IR ، MR, XR, CR, SR , DS ماذا تعني هذه الرموز

NIFLURIL 700 MG, Suppositoire adulte

Antifongiques مضادات الفطريات

Popular posts from this blog

Kana Brax Laberax

TRIPASS XR تري باس

PARANTAL 100 MG, Suppositoire بارانتال 100 مجم تحاميل

الكبد الدهني Fatty Liver

الم اسفل الظهر (الحاد) الذي يظهر بشكل مفاجئ bal-agrisi

SEDALGIC 37.5 MG / 325 MG, Comprimé pelliculé [P] سيدالجيك 37.5 مجم / 325 مجم ، قرص مغلف [P]

نمـو الدمـاغ والتطـور العقـلي لـدى الطفـل

CELEPHI 200 MG, Gélule

أخطر أنواع المخدرات فى العالم و الشرق الاوسط

Archive

Show more