After dipping the strip, check the pH area. If the color on
the pad is not uniform, read the reagent area immediately,
comparing the darkest color to the appropriate Color Chart.
All reagent areas may be read between 1 and 2 minutes for
identifying negative specimens and for determination of
the pH and SG. Color changes that occur after 2 minutes
are of no diagnostic value. If using strips instrumentally,
the instrument will automatically read each reagent area
For best results, performance of reagent strips should
be confirmed by testing known negative and positive
specimens or controls whenever a new bottle is first
opened. Negative and positive specimens or controls
may also be randomly hidden in each batch of specimens
tested. Water should not be used as a negative control.
Each laboratory should establish its own goals for
adequate standards of performance, and should question
handling and testing procedures if these standards are not
met. Chek-stix® Positive and Negative Control Strips, with
positive, negative or defined results, provide a convenient
basis for a urinalysis quality control program.
Because of the various constituents that are added to
commercial controls other than Chek-stix Control Strips,
or the way in which they are processed, specific gravity
values determined using Bayer Reagent Strips may not
always correspond with values given in the product inserts
Due to its specificity for acetoacetic acid, the ketone
reagent area may not react with commercial controls other
than Chek-stix Positive Control Strips. If questionable
results are obtained with the ketone reagent area, strip
reactivity should be checked with Chek-stix Positive
Control Strips or by testing negative and positive clinical
specimens that have been identified as positive or negative
Results with Bayer Reagent Strips are obtained in
clinically meaningful units directly from the Color Chart
comparison when using strips visually. With instrumental
use, the reagent pads are “read” by the instrument and
the results are displayed or printed. The color blocks and
instrumental display values represent nominal values;
actual values will vary around the nominal values.
No laboratory tests, definitive diagnostic or therapeutic
decisions should be based on any single result or method.
Substances that cause abnormal urine color, such as drugs
containing azo dyes (e.g. Pyridium, Azo Gantrisin, Azo
Gantanol), nitrofurantoin (Macrodantin, Furadantin),
and riboflavin, may affect the readability of the reagent
areas on urinalysis reagent strips. The color development on
the reagent pad may be masked, or a color reaction may be
produced on the pad that could be interpreted visually and/
or instrumentally as a false positive.
Ascorbic acid concentrations of 50 mg/dL or greater may
cause false negatives for specimens containing small
amounts of glucose (75–125 mg/dL). Ketone bodies
reduce the sensitivity of the test; moderately high ketone
dL) but the combination of such ketone levels and low
glucose levels is metabolically improbable in screening.
The reactivity of the glucose test decreases as the SG of the
urine increases. Reactivity may also vary with temperature.
Indican (Indoxyl sulfate) can produce a yellow-orange
to red color response that may interfere with the
interpretation of a negative or a positive bilirubin reading.
Metabolites of Lodine (etodolac) may cause false positive
or atypical results; ascorbic acid concentrations of 25 mg/
dL or greater may cause false negatives. Since very small
amounts of bilirubin may be found in the earliest phases
of liver disease, the user must consider whether the
sensitivity of Bayer Reagent Strips to bilirubin is sufficient
False positive results (Trace or less) may occur with highly
pigmented urine specimens or those containing large
amounts of levodopa metabolites. Compounds such as
mesna (2-mercaptoethane sulfonic acid) that contain
sulfhydryl groups may cause false positive results or an
The chemical nature of the Bayer SG test may cause
slightly different results from those obtained with other
specific gravity methods when elevated amounts of
certain urine constituents are present. Highly buffered
alkaline urines may cause low readings relative to other
methods. Elevated specific gravity readings may be
obtained in the presence of moderate quantities (100–750
Elevated specific gravity may reduce the reactivity of
the blood test. Clapoten (captopril) may also cause
decreased reactivity. Certain oxidizing contaminants,
such as hypochlorite, may produce false positive results.
Microbial peroxidase associated with urinary tract
infection may cause a false positive reaction. Levels of
ascorbic acid normally found in urine do not interfere
If proper procedure is not followed and excess urine
remains on the strip, a phenomenon known as “runover”
may occur, in which the acid buffer from the protein
reagent will run onto the pH area, causing a false lowering
False positive results may be obtained with highly buffered
or alkaline urines. Contamination of the urine specimen
with quaternary ammonium compounds (e.g. from
some antiseptics and detergents) or with skin cleansers
containing chlorhexidine may also produce false positive
The reagent area may react with interfering substances
known to react with Ehrlich’s reagent, such as
p-aminosalicylic acid and sulfonamides. Atypical color
reactions may be obtained in the presence of high
concentrations of ρ-aminobenzoic acid. False negative
results may be obtained if formalin is present. Strip
reactivity increases with temperature; the optimum
temperature is 22–26°C. The test is not a reliable method
for the detection of porphobilinogen. The absence of
urobilinogen cannot be determined with this test.
Expected values for the typical “normal” healthy
population and the abnormal population are listed below
for each reagent. Exact agreement between visual results
and instrumental results might not be found because of
the inherent differences between the perception of the
human eye and the optical systems of the instruments.
The kidney normally excretes small amount of glucose.
These amounts are usually below the sensitivity of this test
but on occasion may produce a color between the negative
and the 100 mg/dL color blocks, and that is interpreted
by the instrument as a positive result. Results at the first
positive level may be significantly abnormal if found
Normally no bilirubin is detectable in urine by even
the most sensitive methods. Even trace amounts of
bilirubin are sufficiently abnormal to require further
investigation. Atypical colors (colors that are unlike the
negative or positive color blocks shown on the Color
Chart) may indicate that bilirubin derived bile pigments
are present in the urine sample and may be masking the
bilirubin reaction. These colors may indicate bile pigment
abnormalities and the urine specimen should be tested
Normal urine specimens ordinarily yield negative results
with this reagent. Detectable levels of ketone may occur
in urine during physiological stress conditions such
as fasting, pregnancy and frequent strenuous exercise.
appear in urine in large amounts before serum ketone
Random urines may vary in specific gravity from
1.001–1.035. Twenty-four hour urines from normal adults
with normal diets and normal fluid intake will have a
specific gravity of 1.016–1.022.
The significance of the Trace reaction may vary among
patients, and clinical judgment is required for assessment
in an individual case. Development of green spots (intact
on the reagent area within 60 seconds indicates the need
for further investigation. Blood is often, but not always,
found in the urine of menstruating females. This test is
highly sensitive to hemoglobin and thus complements the
Both the normal and abnormal urinary pH range is from
Normally, no protein is detectable in urine, although
the normal kidney excretes a minute amount. A color
matching any block greater than Trace indicates significant
proteinuria. For urine of high specific gravity, the test area
may most closely match the Trace color block even though
only normal concentrations of protein are present. Clinical
No comments:
Post a Comment
اكتب تعليق حول الموضوع